LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Control Kit Catalog #: L016080109002 (EUA) The LumiraDx SARS-CoV-2 Qu... — FDA record Z-0135-2023

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 23, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Control Kit Catalog #: L016080109002 (EUA) The LumiraDx SARS-CoV-2 Quality Controls are intended for liquid quality control testing performed on the LumiraDx Instrument when used with the LumiraDx SARS-CoV-2 Ag Test Strip. The Quality Controls provide users with assurance that the device is performing within specification
Reason
Assigned vial lot numbers for the incorrect type of product. SARS-CoV-2 & Flu A/B lot numbers were incorrectly assigned to the positive and negative vial lot numbers for SARS-CoV-2 Ag LQCs within batch 2002204017, which should have been assigned SARS-CoV-2 Ag lot numbers resulting in an error message and preventing use of the LQC
Classification
Class III
Recalling firm
LumiraDx
Quantity
83 kits
Agency-reported status
Terminated
Source record identifier
Z-0135-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLQC Kit: 2002204017 Positive Control Vial: 1000 1710 0014 0141 Negative Control Vial: 1000 1720 0014 0142 Exp. Date: 01-Mar-2023

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 2, 2022
Recall initiation date
Sep 23, 2022
Center classification date
Oct 26, 2022
Termination date
Dec 13, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.