Product/Part: MitraClip XTR Clip Delivery System US/CDS0601-XTR, MitraClip G4 Clip Delivery System NT US/CDS0701-NT, ... — FDA record Z-0115-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 8, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Product/Part: MitraClip XTR Clip Delivery System US/CDS0601-XTR, MitraClip G4 Clip Delivery System NT US/CDS0701-NT, MitraClip G4 Clip Delivery System NTW US/CDS0701-NTW, MitraClip G4 Clip Delivery System XT US/CDS0701-XT, MitraClip G4 Clip Delivery System XTW US/CDS0701-XTW. For cardiac procedures.
Reason
An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.
Classification
Class II
Recalling firm
Abbott Vascular
Quantity
18,042
Agency-reported status
Ongoing
Source record identifier
Z-0115-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityPart/UDI: CDS0601-XTR/08717648226366, CDS0701-NT/08717648230943, CDS0701-NTW/08717648230967, CDS0701-XT/08717648230981, CDS0701-XTW/08717648231001, All lots.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT, AU, BE, BG, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK, HU, ID, IE, IL, IN, IR, IT, JO, JP, KR, KW, LB, LU, MT, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RE, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 26, 2022
Recall initiation date
Sep 8, 2022
Center classification date
Oct 20, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.