Product
stryker Tornier Perform Reversed Augmented Glenoid Lateralized Baseplate, 25mm, Offset +3mm, REF DWJ502
Reason
One lot of Tournier Perform Reversed Lateralized Baseplates were found to have been manufactured with 6mm spacers instead of 3mm spacers. This nonconformance prevents the central screw from being able to thread into the baseplate.
Classification
Class II
Recalling firm
Tornier, Inc
Quantity
30 units
Agency-reported status
Ongoing
Source record identifier
Z-0113-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 10846832062017, Lot Number 1756123

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: CA, CO, IL, MA, MN, NC, OR, RI, SC, TN

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 25, 2023
Recall initiation date
Sep 15, 2023
Center classification date
Oct 13, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.