Astral 100 and Astral 150 ventilators — FDA record Z-0111-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 13, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Astral 100 and Astral 150 ventilators
Reason
If ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a Total Power Failure (TPF) alarm should sound, but it's powered by a supercapacitor, which degrades over time, which may cause TPF alarm to sound for less than 2 minutes or not at all
Classification
Class I
Recalling firm
ResMed Ltd.
Quantity
55,279 (US 16,634; OUS 38,645)
Agency-reported status
Ongoing
Source record identifier
Z-0111-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll devices manufactured between 2013 and 2019. Device label lists the serial number, which includes the year of manufacture. For example, if the serial number is 22151234567, the 3rd and 4th digit are 15, meaning the device was manufactured in 2015. Astral 100, UDI: 00619498003259 Astral 150, UDI: 00619498003266

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of CA, NJ, VA, IN, MI, WV, NY, WI, ND, IL, NC, FL, CO, TX, AL, LA, MD, OR, AZ, MN, WA, PA, SC, MA, ID, GA, SD, TN, NH, OH, MS, CT, IA, KY, UT, NE, MO, NV, AR, WY, RI, KS, VT, ME, OK, PR, MO, DC and the countries of Taiwan, Israel, Spain, Slovenia, Belgium, Italy, Australia, Guatemala, Argentina, Uruguay, Chile, France, Martinique, Germany, French Polynesia, Korea, Egypt, Colombia, United Kingdom, Jersey, Ecuador, Reunion, Greece, Brazil, Cyprus, Saudi Arabia, Czech Republic, Thailand, New Caledonia, Mauritius, Canada, Vietnam, Indonesia, Pakistan, Japan, Guadeloupe, South Africa, Iran, Hungary, Tunisia, Nepal, Iceland, New Zealand, Austria, Portugal, Malaysia, Norway, Netherlands, Sweden, Lebanon, Finland, Singapore, United Arab Emirates, Peru, Oman, Philippines, Bahrain, French Guiana, Mexico, Sri Lanka, India, Kuwait, Bangladesh, Kenya, Myanmar, Turkey, Bulgaria, Croatia, Hong Kong, Saint Martin, Qatar, Jordan, Romania, Maldives, Denmark, Poland.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 1, 2023
Recall initiation date
Sep 13, 2023
Center classification date
Oct 23, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.