Sapphire Infusion Pumps, Models: Multi-Therapy (REF 15031-000-0028), Epidural (REF 15032-000-0027) and SapphirePlus (... — FDA record Z-0094-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 11, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Sapphire Infusion Pumps, Models: Multi-Therapy (REF 15031-000-0028), Epidural (REF 15032-000-0027) and SapphirePlus (REF 15038-000-0001)
Reason
Infusion Pumps with affected software revision may fail to detect air in line, which may lead to air embolism.
Classification
Class I
Recalling firm
EITAN MEDICAL LTD
Quantity
1383
Agency-reported status
Ongoing
Source record identifier
Z-0094-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySoftware Revision: 16.10.1 or 16.10.2 REF/UDI-DI: 15031-000-0028/7290109150109, 15032-000-0027/7290109150147, 15038-000-0001/7290109150161

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 25, 2023
Recall initiation date
Sep 11, 2023
Center classification date
Oct 19, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.