CARESITE SMALLBORE EXT SET T-PORT 5 IN., Product Code 470114, 100 units per carton. The A6 Access Device is a valve i... — FDA record Z-0092-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 15, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
CARESITE SMALLBORE EXT SET T-PORT 5 IN., Product Code 470114, 100 units per carton. The A6 Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood.
Reason
The extension set has a high probability of potential failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.
Classification
Class II
Recalling firm
B. Braun Medical, Inc.
Quantity
354,200 US; 1,800 OUS
Agency-reported status
Ongoing
Source record identifier
Z-0092-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI (GTIN): 04046964298691 Lots/Expiration Dates 61762215 15-Dec-23 61762791 22-Jan-24 61767369 26-Feb-24 61779625 7-Jun-24 61791895 11-Aug-24 61806047 4-Nov-24

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Distribution US nationwide, Canada and Saudi Arabia.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 26, 2022
Recall initiation date
Sep 15, 2022
Center classification date
Oct 14, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records