Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples — FDA record Z-0085-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 4, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples
Reason
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
Classification
Class II
Recalling firm
LEICA BIOSYSTEMS NUSSLOCH GMBH
Quantity
1 unit
Agency-reported status
Ongoing
Source record identifier
Z-0085-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 04049188066237, Serial Number 990

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 23, 2024
Recall initiation date
Sep 4, 2024
Center classification date
Oct 15, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records