Clip COVID Rapid Antigen Test Kit (25-pack), Catalog Number: MDL002 — FDA record Z-0080-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 2, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Clip COVID Rapid Antigen Test Kit (25-pack), Catalog Number: MDL002
Reason
COVID test kit shelf life was periodically extended, as real-time stability test timepoints passed, without FDA concurrence.
Classification
Class II
Recalling firm
LUMINOSTICS, INC dba CLIP HEALTH
Quantity
10037 boxes of 25
Agency-reported status
Ongoing
Source record identifier
Z-0080-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLots: CC-10094-A, CC-10115-A, CC-10123-A, CC-10129-A, CC-10136-A, CC-10144-A, CC-10148-A, CC-10166-A, CC-10168-A, CC-10168-B, CC-10189-A, CC-10193-A, CC-10300-A, CC-10328-A, CC-10329-A, CC-10330-A, CC-10361-A, CC-10364-A, CC-10365-A, CC-10368-A, CC-10369-A, CC-10395-A, CC-10396-A, CC-10397-A, CC-10398-A, CC-10399-A, CC-10439-A, CC-10440-A, CC-10440-B, CC-10441-A, CC-10441-B, CC-10609-A, CC-10609-B, CC-10611-A, CC-10613-A, CC-10615-A, CC-10654-A, CC-10667-A, CC-10710-A, CC-10710-B, CC-10712-A, CC-10714-A, CC-10714-B, CC-10725-A, CC-10727-A, CC-10729-A, CC-10766-A, CC-10768-A, CC-10809-S, CC-10811-S, CC-10874-A, CC-10874-B, CC-10874-C, CC-50002-S, CC-50003-S, CC-50004-S Expiration Dates: expiry.clipcovid.com

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US nationwide distribution

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 19, 2022
Recall initiation date
Sep 2, 2022
Center classification date
Oct 12, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records