Olympus High-Flow Insufflation Unit, Model UHI-4. For laparoscopic surgery. — FDA record Z-0075-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 22, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Olympus High-Flow Insufflation Unit, Model UHI-4. For laparoscopic surgery.
Reason
There have been reports of patients suffering arrhythmias, reported as short cardiac arrests, during surgical procedures where UHI-4s were used. These events may have been due to an over insufflation of the abdominal cavity resulting from use of the UHI-4 during the procedures.
Classification
Class I
Recalling firm
Olympus Corporation of the Americas
Quantity
3136 units
Agency-reported status
Ongoing
Source record identifier
Z-0075-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 04953170324147; All Serial Numbers

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 25, 2023
Recall initiation date
Sep 22, 2023
Center classification date
Oct 18, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.