Recall archive / FDA device enforcement
1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of... — FDA record Z-0063-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Sep 5, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- 1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of bone, bone cement and teeth in a variety of surgical procedures including but not limited to dental, ENT (ear, nose and throat), neuro, spine and endoscopic applications Catalog Number: 5820-066-012
- Reason
- May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only occur if there is contact with cerebrospinal fluid and thus, neurosurgical/spinal procedures may be considered a population at greatest risk.
- Classification
- Class II
- Recalling firm
- Stryker Corporation
- Quantity
- 307 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0063-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | GTIN: 04546540362346 Lot number: 22329017 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Columbia.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Oct 23, 2024
- Recall initiation date
- Sep 5, 2024
- Center classification date
- Oct 11, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.