Enterprise Imaging XERO Viewer- software application used for reference and diagnostic viewing of multispecialty medi... — FDA record Z-0055-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 6, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Enterprise Imaging XERO Viewer- software application used for reference and diagnostic viewing of multispecialty medical imaging and non-imaging data . Enterprise Imaging XERO Viewer 8.0.0, 8.0.1, 8.1, 8.1.1, 8.1.2, 8.1.3,8.1.4, 8.2.0.000, 8.2.1.000, 8.2.2.000, 8.3.0.000, 8.3.1.000, 8.3.2.000, 8.4.0.000, 8.4.1.000 (including all SU s and HF s)
Reason
Software defect, when an image has non-square pixel spacing, the reference/localizer line is incorrectly placed on the scout image in the XERO Viewet, may display the incorrect crosshair/line position on the non-coplanar/scout image with respect to the axial slice location. Inaccurate placement of reference lines or crosshairs have potential risk of misdiagnosis or mistreatment due to unintended shifts in the display of anatomical locations.
Classification
Class II
Recalling firm
Agfa Healthcare NV
Quantity
801 units
Agency-reported status
Ongoing
Source record identifier
Z-0055-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 05400874000710 Software Versions affected: All versions of Enterprise Imaging XERO Viewer are affected 8.1.2 or higher 8.1.4 or higher 8.2.0.100 or higher 8.2.2.000 or higher 8.3.2.000 or higher 8.4.0.000 or higher Enterprise Imaging XERO Viewer main version streams 8.4.0, 8.4.1, 8.3.2, 8.3.1, 8.3.0, 8.2.2, 8.2.1, 8.2.0, 8.1.5, 8.1.4, 8.1.3, 8.1.2, 8.1.1, 8.1, 8.0.1, 8.0.0, ICIS View 2014.1, ICIS View 2014.2

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution. Foreign: Argentina, Australia, Belgium, Chile, Canada, Colombia, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Georgia, Greece, Hong Kong, Iceland, Ireland, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Luxembourg, Mexico, Netherlands, Nicaragua, Norway, Oman, Pakistan, Peru, Poland, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Tanzania, Tunisia, United Arab Emirates, United Kingdom

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 16, 2024
Recall initiation date
Sep 6, 2024
Center classification date
Oct 10, 2024