Check-Flo Performer Introducer Set, Reference Part Number RCF-5.0-35-45-J, Order Number G27242. Performer Introducers... — FDA record Z-0045-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 16, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Check-Flo Performer Introducer Set, Reference Part Number RCF-5.0-35-45-J, Order Number G27242. Performer Introducers are designed to perform as a guiding sheath and/or introducer sheath and come with a Check Flo valve.
Reason
Cook Medical is conducting removal of a lot of Check-Flo Performer Introducer Set due to complaints that the white caps that connect the shaft to the hub of these products may disconnect.
Classification
Class II
Recalling firm
Cook Incorporated
Quantity
50 units
Agency-reported status
Ongoing
Source record identifier
Z-0045-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00827002272422; Lot Number: 15506769

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

International distribution to the country of Australia.

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Oct 16, 2024
Recall initiation date
Sep 16, 2024
Center classification date
Oct 9, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.