The EPIONE device v1.0.2 is a user controlled, stereotactic accessory device intended to assist in the planning and m... — FDA record Z-0037-2025

Reported completed

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 5, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
The EPIONE device v1.0.2 is a user controlled, stereotactic accessory device intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures. During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument placement and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data.
Reason
Quantum Surgical has become aware that a component (central axis) inside the Needle Guide product (product number 02-0027) is the subject of rust.
Classification
Class II
Recalling firm
QUANTUM SURGICAL SAS
Quantity
2 units
Agency-reported status
Completed
Source record identifier
Z-0037-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 37603054002NGXL, serial numbers: 0321015 and 0321017

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the state of FL.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 16, 2024
Recall initiation date
Aug 5, 2024
Center classification date
Oct 9, 2024