Recall archive / FDA device enforcement
BD Alaris Pump Infusion Sets for the following reference numbers: C24101E, Amber Tubing, 15 Micron Filter, 0.2 Micron... — FDA record Z-0035-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Sep 11, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- BD Alaris Pump Infusion Sets for the following reference numbers: C24101E, Amber Tubing, 15 Micron Filter, 0.2 Micron Filter, SmartSite Y-Site 10015414, 180 Micron Filter 2447-0007, Ball Valve SmartSite Port (Burette) 2 SmartSites Y-Sites 24201-0007, 2 SmartSite Y-sites 10012645, Back Check Valve, Non-vented, SmartSite Y-site 2426-0500, Back Check Valve, 3 SmartSite Y-sites 2450-0500, Check Valve 2 Ganged 3-Way Stopcocks 3 SmartSite Y-sites 11522558, Ball Valve, Back Check Valve, 3 SmartSite Y-sites 2410-0500, 2 SmartSite Y-Sites 2232-0007, 0.2 Micron Filter 11613191, 0.2 Micron Filter, Ball Valve, SmartSite" Port (Burette), 2 SmartSite Y-sites 10012144, 3 SmartSite Y-Sites, Check Valve 10013034, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 4-way Stopcock 10013037, Non-Vented, 180 Micron Filter, Low Sorbing Tubing Segment, 1 SmartSite, Y-Site 11582773, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 3-way Stopcock
- Reason
- Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
- Classification
- Class I
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- Not supplied
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0035-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | All Lots/ UDI: C24101E - 10885403235528, 10015414 - 10885403233999, 2447-0007 - 10885403235078, 24201-0007 - 10885403476341, 10012645 - 10885403233753, 2426-0500 - 7613203020992, 11522558 - 10885403232305, 2410-0500 - 7613203012492, 2232-0007 - 10885403463969, 11613191 - 10885403276026, 10012144 - 7613203021159, 10013034 - 10885403232312, 10013037 - 10885403233784, 11582773 - 10885403239588 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US: Nationwide OUS: United Arab Emirates Australia Belgium Bahrain Botswana Canada Switzerland China France United Kingdom Gibraltar Israel India Kuwait Mexico Malaysia Netherlands New Zealand Philippines Pakistan Qatar Saudi Arabia Singapore Taiwan, Province of China South Africa
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Oct 15, 2025
- Recall initiation date
- Sep 11, 2025
- Center classification date
- Oct 8, 2025
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.