REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx O... — FDA record Z-0025-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 29, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx Only, Sterile EO
Reason
There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed.
Classification
Class II
Recalling firm
Becton Dickinson Infusion Therapy Systems Inc.
Quantity
98,000 units
Agency-reported status
Ongoing
Source record identifier
Z-0025-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot # 2238464/ UDI-DI: (01)00382903868629

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of CA, FL, GA, MA, MN, MO, MS, NJ, OH, TX.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 11, 2023
Recall initiation date
Aug 29, 2023
Center classification date
Oct 3, 2023