Recall archive / FDA device enforcement
ENT RES, 100 ML, Non-Sterile W CLAMP & FEMALE LUER, Product Code 68-2231-28 — FDA record Z-0018-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jun 26, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- ENT RES, 100 ML, Non-Sterile W CLAMP & FEMALE LUER, Product Code 68-2231-28
- Reason
- Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
- Classification
- Class II
- Recalling firm
- Smiths Medical ASD Inc.
- Quantity
- 120960 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0018-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI/DI 30610586044186, Lot Numbers: 3997242-ENG, 4040002AB, 4040000AB, 4040001AB, 4040008AB, 4040007AB, 4040009AB, 4146021AB |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution.
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Oct 30, 2024
- Recall initiation date
- Jun 26, 2024
- Center classification date
- Oct 18, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.