CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24JP — FDA record Z-0015-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 26, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24JP
Reason
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Classification
Class II
Recalling firm
Smiths Medical ASD Inc.
Quantity
109563 units
Agency-reported status
Ongoing
Source record identifier
Z-0015-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUID/DI NA, Lot/Serial Numbers: 3983315, 3971709, 3983316, 3983320, 3983329, 4013361, 4013363, 4013373, 4013377, 4013378, 4019427, 4019429, 4034028, 4037753, 4052410, 4040250, 4042843, 4048894, 4046846, 4052412, 4052413

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution.

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Oct 30, 2024
Recall initiation date
Jun 26, 2024
Center classification date
Oct 18, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records