Recall archive / FDA device enforcement
BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System (Base),... — FDA record Z-0009-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Aug 12, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System (Base), REF: 663128, 662588, used with software V 1.4.1. Intended use: to prepare human specimens to be analyzed.
- Reason
- Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute counts in tube 1 than tube 2; could result in additional blood collection, delayed results/patient therapy.
- Classification
- Class II
- Recalling firm
- Becton, Dickinson and Company, BD Bio Sciences
- Quantity
- 130
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0009-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | REF/UDI-DI/Serial Numbers: 666340/382906663405/9230010101, 9230010102, 9230010103, 9230010104, 9230010105, 9230010107, 9230010108, 9230010109, 9230010112, 9230010113, 9230010114, 9230010115, 9230010116, 9230010118, 9230010119, 9230010120. 666339/9230000102, 9230000103, 9230000104, 9230000105, 9230000106, 9230000112, 9230000113, 9230000114, 9230000118, 9230000120, 9230000121, 9230000123, 9230000124, 9230000126, 9230000128, 9230000130, 9230000132, 9230000134, 9230000135, 9230000138, 9230000144, 9230000145, 9230000131, 9230000152, 9230000149, 9230000030, 9230000101, 9230000108, 9230000109, 9230000110, 9230000111, 9230000115, 9230000116, 9230000117, 9230000119, 9230000122, 9230000125, 9230000127, 9230000129, 9230000136, 9230000139, 9230000141, 9230000142, 9230000143. 663128/382906631282/9220030004, 9220030005, 9220030008, 9220030016, 9220030017, 9220030027, 9220030029, 9220030031, 9220030043, 9220030053, 9220030057. 662588/9220010119, 9220010179, 9220010182, 9220010232, 9220010236, 9220010250, 9220010265, 9220010275, 9220010293, 9220010311, 9220010313, 9220010316, 9220010324, 9220010327, 9220010335, 9220010338, 9220010341, 9220010344, 9220010348, 9220010349, 9220010292, 9220010112, 9220010120, 9220010125, 9220010127, 9220010128, 9220010130, 9220010133, 9220010138, 9220010154, 9220010161, 9220010183, 9220010188, 9220010197, 9220010215, 9220010225, 9220010231, 9220010238, 9220010240, 9220010242, 9220010244, 9220010256, 9220010257, 9220010260, 9220010263, 9220010268, 9220010269, 9220010279, 9220010280, 9220010289, 9220010290, 9220010294, 9220010296, 9220010329, 9220010332, 9220010345, 9220010354, 9230000031, 99220010292. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US: NJ, UT, WI, TX, ID, AK, MN, MA, MD, TN, MI, CA, VA OUS: Ireland, Austria, Italy, Germany, Belgium, Finland, Canada, France, Spain, New Zealand, United Kingdom, Japan, Australia, Norway, Singapore, Sweden, Turkey, Brazil, Colombia, Netherlands, United Arab Emirates, Saudi Arabia, Mexico, Switzerland, Costa Rica, China, Azerbaijan, Algeria, Jordan, India
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- Oct 9, 2024
- Recall initiation date
- Aug 12, 2024
- Center classification date
- Oct 3, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.