Proboscis/Piston Assembly, Part Number J55375, for VITROS 5,1 FS Chemistry System (Product Code 6801375), VITROS 5,1 ... — FDA record Z-0008-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 9, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Proboscis/Piston Assembly, Part Number J55375, for VITROS 5,1 FS Chemistry System (Product Code 6801375), VITROS 5,1 Chemistry System (Refurbished) (Product Code 6801890), VITROS 3600 Immunodiagnostic System (Product Code 6802783), VITROS 3600 Immunodiagnostic System (Refurbished) (Product Code 6802914), VITROS 4600 Chemistry System (Product Code 6802445), VITROS 4600 Chemistry System (Refurbished) (Product Code 6900440), VITROS 5600 Integrated System (Product Code 6802413), VITROS 5600 Integrated System (Refurbished) (Product Code 6802915), VITROS XT 3400 Chemistry System (Product Code 6844458), and VITROS XT 7600 Integrated System (Product Code 6844461)
Reason
Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as intended due to an out-of-specification barrel component. This may cause VITROS VersaTips and/or VITROS MicroTips to be improperly seated on the proboscis, which may result in falsely elevated Gentamicin and Valproic Acid assay results as well as an increased frequency of condition codes resulting in delayed patient results. The issue was identified by internal testing when reported instruments failed to pass release testing and were traced to the molded Proboscis part J56639 with date codes from 040722 onward.
Classification
Class II
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
9,860 assemblies
Agency-reported status
Ongoing
Source record identifier
Z-0008-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAssemblies with a date code between 040722 - 121922

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 11, 2023
Recall initiation date
Aug 9, 2023
Center classification date
Oct 2, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.