Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001 — FDA record Z-0008-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 30, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001
Reason
Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate an alternative form of fixation. Additionally, there is a potential for adverse tissue reaction if the surgeon attempts to use the guide regardless of tight fit and plastic debris is created and ends up in the wound
Classification
Class II
Recalling firm
Synthes (USA) Products LLC
Quantity
191 units
Agency-reported status
Ongoing
Source record identifier
Z-0008-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00810633022047 Lot Numbers: MHC220119, MHC220120 Expiry Date: 1 June 2027

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 12, 2022
Recall initiation date
Aug 30, 2022
Center classification date
Oct 6, 2022