Zimmer, Inc. recall records

70 reported ongoing

71 records in this archive. Latest source date May 20, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609420 Software Version: N/A Product Description: Aff...

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunio...

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.

Official source for Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nail...
FDA device enforcement

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609440 Software Version: N/A Product Description: Aff...

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunio...

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.

Official source for Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nail...
FDA device enforcement

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815809280 Software Version: N/A Product Description: Aff...

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunio...

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.

Official source for Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nail...
FDA device enforcement

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509300 Software Version: N/A Product Description: Aff...

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunio...

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.

Official source for Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nail...
FDA device enforcement

Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, Bearing, Model/Catalog Number: 110031012; semi constrained hip prosthesis

Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E¿ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.

Class IIReported ongoing

Distribution: US: OH, GA, WI, NE, IA, IL, CA, NY, MD, NC, PA, AL, TN, MI, WA, FL, TX, KS, NM, AZ, and CANADA, AUSTRALIA, NETHERLANDS

Official source for Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Si...
FDA device enforcement

Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 46 mm O.D., Size G, Bearing, Model/Catalog Number: 110031013; semi constrained hip prosthesis

Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E¿ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.

Class IIReported ongoing

Distribution: US: OH, GA, WI, NE, IA, IL, CA, NY, MD, NC, PA, AL, TN, MI, WA, FL, TX, KS, NM, AZ, and CANADA, AUSTRALIA, NETHERLANDS

Official source for Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 46 mm O.D., Si...
FDA device enforcement

Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Un...

The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.

Official source for Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prost...
FDA device enforcement

Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +3 Neck, Model Number 802602203; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Un...

The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.

Official source for Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +3 Neck, Model Number 802602203; Prost...
FDA device enforcement

Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, -3 Neck, Model Number 802802201; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented

The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.

Official source for Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, -3 Neck, Model Number 802802201; Prosthesis...
FDA device enforcement

Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +0 Neck, Model Number 802802202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented

The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.

Official source for Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +0 Neck, Model Number 802802202; Prosthesis...
FDA device enforcement

Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +3 Neck, Model Number 802802203; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented

The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.

Official source for Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +3 Neck, Model Number 802802203; Prosthesis...
FDA device enforcement

NexGen LPS Flex, Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis, Model Number 00-5962-042-10

The "Use with plate 7, 8, 9, 10" compatibility statement on top of the label does not match with the compatibility statements "USE WITH PLATE 5, 6" in the translations section of the same label.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states pf CA, KS, MA, MD, ME, NE, NJ, OR, SD, and TX. The countries of Canada, China, India, Netherlands, Singapore.

Official source for NexGen LPS Flex, Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented p...
FDA device enforcement

NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Purple/C,D - Height 14 mm, Item Number 00-5994-022-14

It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of CA, FL, GA, KS, MO, NE, NJ, PA, TN and the countries of CHINA, INDIA, KOREA, MALAYSIA, NETHERLANDS, SINGAPORE, TAIWAN.

Official source for NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Purple/C,D - Height 14 mm, Item...
FDA device enforcement

NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10

It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of CA, FL, GA, KS, MO, NE, NJ, PA, TN and the countries of CHINA, INDIA, KOREA, MALAYSIA, NETHERLANDS, SINGAPORE, TAIWAN.

Official source for NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item...
FDA device enforcement

Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patien...

A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, re...

Class IIReported ongoing

Distribution: Nationwide

Official source for Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive sur...
FDA device enforcement

CPT Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, used in Hip Arthroplasty; Item Numbers 00811400000 00811400010 00811400100 00811400110 00811400200 00811400210 ...

Affected product has an increased risk of postoperative perisprosthetic femoral fracture (PFF). The IFU is being updated to reflect the risk of PFF.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution.

Official source for CPT Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, used in Hip Arthroplasty; Item ...
FDA device enforcement

NexGen CR-Flex Femoral Component, Size E and Size E Minus (E-). Component of the NexGen CR-Flex Porous Femur System.

Out of specification violation of devices that results in a gap existing within the fixture that presses the fiber metal pad to the substrate, causing the pad to not fully bond to the substrate.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide including the states of AK, AL, CA, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA...

Official source for NexGen CR-Flex Femoral Component, Size E and Size E Minus (E-). Component of the NexGen CR-Flex P...
FDA device enforcement

ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 4 Holes, 80 MM, Sterile- Indicated for temporary internal fixation and stabilization of ost...

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.

Official source for ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 4 Holes, 80...
FDA device enforcement

ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Sterile- Indicated for temporary internal fixation and stabilization of os...

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.

Official source for ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 6 Holes, 10...

Reasons in these listed records

  • Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage ...
  • Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E¿ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
  • The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
  • The "Use with plate 7, 8, 9, 10" compatibility statement on top of the label does not match with the compatibility statements "USE WITH PLATE 5, 6" in the translations section of the same label.

These examples come from this result page and are not a ranking of every recall reason.

Checking Zimmer, Inc.

Is there a recall on zimmer, inc. right now?

This archive contains 71 records, including 70 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking