TELEFLEX LLC recall records

139 reported ongoing

140 records in this archive. Latest source date May 7, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or tissue ligation

Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the c...

Class IIReported ongoing

Distribution: worldwide

Official source for Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgica...
FDA device enforcement

Teleflex (manufacturer: Vascular SOLUTIONS) D-Stat Flowable Hemostat, REF 4000; for sealing residual oozing of tissue tracts of femoral access sites that have been previously cl...

The product contains an expired diluent vial. If the impacted product is used, the sterility of the diluent component cannot be assured as it has passed beyond the studied shelf life of the product.

Class IIReported ongoing

Distribution: US Nationwide

Official source for Teleflex (manufacturer: Vascular SOLUTIONS) D-Stat Flowable Hemostat, REF 4000; for sealing resid...
FDA device enforcement

Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal. Intended for patients requiring endotracheal intubation

Potential for pilot balloon non-inflation or cuff non-deflation.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OH, OK,...

Official source for Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal. Intended for patients requiring en...
FDA device enforcement

Endotracheal Tube labeled as the following: a. Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal; b. Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressu...

Potential for pilot balloon non-inflation or cuff non-deflation.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OH, OK,...

Official source for Endotracheal Tube labeled as the following: a. Flexi-Set Cuffed Endotracheal Tube and Stylet Set,...
FDA device enforcement

Endotracheal Tube labeled as the following: a. Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff; b. Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low ...

Potential for pilot balloon non-inflation or cuff non-deflation.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OH, OK,...

Official source for Endotracheal Tube labeled as the following: a. Endotracheal Tube oral/nasal Magill, High Volume, ...

Reasons in these listed records

  • Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
  • There are incomplete pre-cleaning instructions within the Instructions for Use (IFU).
  • Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a ...
  • The product contains an expired diluent vial. If the impacted product is used, the sterility of the diluent component cannot be assured as it has passed beyond the studied shelf life of the product.

These examples come from this result page and are not a ranking of every recall reason.

Checking TELEFLEX LLC

Is there a recall on teleflex llc right now?

This archive contains 140 records, including 139 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking