FDA device enforcement

Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL

There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the mi...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries ...

Official source for Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL
FDA device enforcement

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position whe...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.

Official source for Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Reasons in these listed records

  • There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y moto...
  • There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops a...

These examples come from this result page and are not a ranking of every recall reason.

Checking Shimadzu Corp. Analytical & Measuring Instruments Div.

Is there a recall on shimadzu corp. analytical & measuring instruments div. right now?

This archive contains 2 records, including 2 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking