Richard Wolf GmbH recall records

5 reported ongoing

5 records in this archive. Latest source date Apr 9, 2025.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

PiezoWave 2 Control Unit, Product Number 100506US, compact myofascial acoustic compression therapy device that enables treatment with piezoelectric shockwaves (Therapeutic massa...

The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat ...

Class IIReported ongoing

Distribution: US State: GA

Official source for PiezoWave 2 Control Unit, Product Number 100506US, compact myofascial acoustic compression therap...
FDA device enforcement

FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 828.051

Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of United Arab Emirates (AE), Argentina (AR), Austria (AT), Belgium (BE), BO, Brazil (BR), Canada (CA), Chile (CL), Czec...

Official source for FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 828.051
FDA device enforcement

FLEX. BIOPSY FORCEPS 5FR WL 550MM, Product Number 829.051

Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of United Arab Emirates (AE), Argentina (AR), Austria (AT), Belgium (BE), BO, Brazil (BR), Canada (CA), Chile (CL), Czec...

Official source for FLEX. BIOPSY FORCEPS 5FR WL 550MM, Product Number 829.051
FDA device enforcement

FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 8735.685

Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of United Arab Emirates (AE), Argentina (AR), Austria (AT), Belgium (BE), BO, Brazil (BR), Canada (CA), Chile (CL), Czec...

Official source for FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 8735.685
FDA device enforcement

FLEX. GRASP. FORCEPS 6.6FR WL 550MM, Product Number 8736.685

Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of United Arab Emirates (AE), Argentina (AR), Austria (AT), Belgium (BE), BO, Brazil (BR), Canada (CA), Chile (CL), Czec...

Official source for FLEX. GRASP. FORCEPS 6.6FR WL 550MM, Product Number 8736.685

Reasons in these listed records

  • The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat when an internal cap...
  • Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.

These examples come from this result page and are not a ranking of every recall reason.

Checking Richard Wolf GmbH

Is there a recall on richard wolf gmbh right now?

This archive contains 5 records, including 5 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.