Respironics California, LLC recall records

5 reported ongoing

5 records in this archive. Latest source date Feb 28, 2022.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Philips Respironics V60 Ventilator (All Models, including: 1053613, 1053614, 1053615, 1053616, 1053617, 1053618, 1076709, 1076716, 1076717, 1137292, DU1053617, U1053617, R105361...

Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.

Class IReported ongoing

Distribution: Worldwide distribution US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND...

Official source for Philips Respironics V60 Ventilator (All Models, including: 1053613, 1053614, 1053615, 1053616, 10...
FDA device enforcement

Philips Respironics V60 Plus Ventilator (All Models, including: 1138747, 1137276)

Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.

Class IReported ongoing

Distribution: Worldwide distribution US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND...

Official source for Philips Respironics V60 Plus Ventilator (All Models, including: 1138747, 1137276)
FDA device enforcement

Philips Respironics V680 Ventilator (All Models, including: P/N 1103044 (850011)), *Distributed outside the United States

Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.

Class IReported ongoing

Distribution: Worldwide distribution US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND...

Official source for Philips Respironics V680 Ventilator (All Models, including: P/N 1103044 (850011)), *Distributed o...
FDA device enforcement

Philips Respironics V60 Ventilator Part Number 1053617

A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may ei...

Class IReported ongoing

Distribution: U.S.: AR, CA, FL, GA, IN, LA, MD, MO, NY, PA, SC, TN and TX. O.U.S. N/A

Official source for Philips Respironics V60 Ventilator Part Number 1053617
FDA device enforcement

Philips Respironics V60 Plus Ventilator Part Number 1138747

A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may ei...

Class IReported ongoing

Distribution: U.S.: AR, CA, FL, GA, IN, LA, MD, MO, NY, PA, SC, TN and TX. O.U.S. N/A

Official source for Philips Respironics V60 Plus Ventilator Part Number 1138747

Reasons in these listed records

  • Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.
  • A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may either activate both v...

These examples come from this result page and are not a ranking of every recall reason.

Checking Respironics California, LLC

Is there a recall on respironics california, llc right now?

This archive contains 5 records, including 5 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking