Qiagen GmbH recall records

4 reported ongoing

4 records in this archive. Latest source date Aug 22, 2023.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

therascreen ¿KRAS RGQ PCR Kit (24)-IVD qualitative PCR assay used on the Rotor-Gene Q MDx instrument for the detection of seven somatic mutations in the human KRAS oncogene REF ...

Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with th...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, FL, IL, IN, KS, MA, NC,NY,OR, PA, TN, TX and the countries of KR, CN

Official source for therascreen ¿KRAS RGQ PCR Kit (24)-IVD qualitative PCR assay used on the Rotor-Gene Q MDx instrum...
FDA device enforcement

QIAstat-Dx Respiratory SARS-CoV-2 Panel -US IVD qualitative test intended for analyzing nasopharyngeal swab (NPS) samples from patients suspected of respiratory infection for th...

Identified a decreased performance reliability rate, Run abortions could cause delayed diagnosis since sample testing would have to be repeated

Class IIReported ongoing

Distribution: DC, AL, CA, FL, GA, ME, NC, NY, OR, TX Foreign: DE, AE, AT, BE, BG, CH, CL, CY, CZ, DE, DK, ES, FI, FR, GB, GR, GT, HK, HU, ID, IN, IT, JO, MC, MU, MY, NL, NA, PL, PT, RO, SE, S...

Official source for QIAstat-Dx Respiratory SARS-CoV-2 Panel -US IVD qualitative test intended for analyzing nasophary...
FDA device enforcement

QIAstat-Dx Respiratory SARS-CoV-2 Panel (V1 and V2, CE IVD, not US IVD, not sold in the US.) Qualitative test intended for analyzing nasopharyngeal swab (NPS) samples from patie...

Identified a decreased performance reliability rate, Run abortions could cause delayed diagnosis since sample testing would have to be repeated

Class IIReported ongoing

Distribution: DC, AL, CA, FL, GA, ME, NC, NY, OR, TX Foreign: DE, AE, AT, BE, BG, CH, CL, CY, CZ, DE, DK, ES, FI, FR, GB, GR, GT, HK, HU, ID, IN, IT, JO, MC, MU, MY, NL, NA, PL, PT, RO, SE, S...

Official source for QIAstat-Dx Respiratory SARS-CoV-2 Panel (V1 and V2, CE IVD, not US IVD, not sold in the US.) Qual...
FDA device enforcement

QIAstat-Dx¿ Respiratory SARS-CoV-2 Panel -IVD intended for the detection and differentiation of nucleic acid from SARS-CoV-2 and the following organism types and subtypes: Adeno...

If cartridges of the affected Serial numbers (SNs) are used, in most cases, the error codes 0xY0003EA and 0xY00045A (where Y is a number between 1 and 4) that lead to a run abortion will occur and no results will be reported

Class IIReported ongoing

Distribution: AL, AR, CA, DC, FL, GA, ID, KS, KY, MD, MN, NJ, NY,PA, SC, TX, VA

Official source for QIAstat-Dx¿ Respiratory SARS-CoV-2 Panel -IVD intended for the detection and differentiation of n...

Reasons in these listed records

  • Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with this flag should be in...
  • Identified a decreased performance reliability rate, Run abortions could cause delayed diagnosis since sample testing would have to be repeated
  • If cartridges of the affected Serial numbers (SNs) are used, in most cases, the error codes 0xY0003EA and 0xY00045A (where Y is a number between 1 and 4) that lead to a run abortion will occur and no results will be reported

These examples come from this result page and are not a ranking of every recall reason.

Checking Qiagen GmbH

Is there a recall on qiagen gmbh right now?

This archive contains 4 records, including 4 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking