Philips North America recall records

126 reported ongoing

126 records in this archive. Latest source date Jul 31, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Philips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, ...

Official source for Philips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.
FDA device enforcement

Patient Information Center iX; Software Version Number: 4.5.0

A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Canada, Czech Republic, Dominican Rep, Finland, France, Germany, Greece, Hungary, India, It...

Official source for Patient Information Center iX; Software Version Number: 4.5.0
FDA device enforcement

Evolution Upgrade 1.5T. Product Codes (REF): (1) 782148, (2) 782116. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for Evolution Upgrade 1.5T. Product Codes (REF): (1) 782148, (2) 782116. MR systems with SW version R...
FDA device enforcement

Evolution Upgrade 3.0T. Product Code (REF): 782143. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for Evolution Upgrade 3.0T. Product Code (REF): 782143. MR systems with SW version R11.1 and R12.1.
FDA device enforcement

Ingenia 1.5T. Product Code (REF): (1) 781341, (2) 781396. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for Ingenia 1.5T. Product Code (REF): (1) 781341, (2) 781396. MR systems with SW version R11.1 and R1...
FDA device enforcement

Ingenia 3.0T. Product Code (REF): (1) 781342, (2) 781377. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for Ingenia 3.0T. Product Code (REF): (1) 781342, (2) 781377. MR systems with SW version R11.1 and R1...
FDA device enforcement

Ingenia 3.0T CX. Product Code (REF): 781271. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for Ingenia 3.0T CX. Product Code (REF): 781271. MR systems with SW version R11.1 and R12.1.
FDA device enforcement

Ingenia Ambition S. Product Code (REF): 782108. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for Ingenia Ambition S. Product Code (REF): 782108. MR systems with SW version R11.1 and R12.1.
FDA device enforcement

Ingenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for Ingenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW ver...
FDA device enforcement

Ingenia Elition S. Product Code (REF): (1) 781357, (2) 782106, (3)782137. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for Ingenia Elition S. Product Code (REF): (1) 781357, (2) 782106, (3)782137. MR systems with SW vers...
FDA device enforcement

Ingenia Elition X. Product Code (REF): (1) 781358, (2) 782107, (3) 782136. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for Ingenia Elition X. Product Code (REF): (1) 781358, (2) 782107, (3) 782136. MR systems with SW ver...
FDA device enforcement

MR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for MR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1.
FDA device enforcement

Upgrade to MR 7700. Product Code (REF): 782130. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for Upgrade to MR 7700. Product Code (REF): 782130. MR systems with SW version R11.1 and R12.1.
FDA device enforcement

SmartPath to dStream for 3.0T. Product Code (REF): 782145. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for SmartPath to dStream for 3.0T. Product Code (REF): 782145. MR systems with SW version R11.1 and R...
FDA device enforcement

SmartPath to dStream for XR and 3.0T. Product Code (REF): 781270. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denm...

Official source for SmartPath to dStream for XR and 3.0T. Product Code (REF): 781270. MR systems with SW version R11....
FDA device enforcement

Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM

Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluo...

Class IIReported ongoing

Distribution: U.S.

Official source for Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XR...
FDA device enforcement

Achieva 3.0T; Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345;

The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Belize, Boli...

Official source for Achieva 3.0T; Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345;
FDA device enforcement

Achieva XR; Model Numbers (REF): (1) 781153, (2) 781253;

The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Belize, Boli...

Official source for Achieva XR; Model Numbers (REF): (1) 781153, (2) 781253;

Reasons in these listed records

  • The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
  • A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.
  • The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
  • Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.

These examples come from this result page and are not a ranking of every recall reason.

Checking Philips North America

Is there a recall on philips north america right now?

This archive contains 126 records, including 126 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking