PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recall records

208 reported ongoing

208 records in this archive. Latest source date Jul 16, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Azurion 3 M15 System Model Numbers: (1) 722064, (2) 722222, (3) 722230, (4)722280(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belg...

Official source for Azurion 3 M15 System Model Numbers: (1) 722064, (2) 722222, (3) 722230, (4)722280(OUS only); Soft...
FDA device enforcement

Azurion 7 B12 System Model Numbers:(1) 722067, (2) 722225, (3) 722235; Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belg...

Official source for Azurion 7 B12 System Model Numbers:(1) 722067, (2) 722225, (3) 722235; Software Version Number - ...
FDA device enforcement

Azurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belg...

Official source for Azurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - Al...
FDA device enforcement

Azurion 7 M12 System Model Numbers: (1)722078, (2)722223, (3)722233; Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belg...

Official source for Azurion 7 M12 System Model Numbers: (1)722078, (2)722223, (3)722233; Software Version Number - Al...
FDA device enforcement

Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belg...

Official source for Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Softwar...
FDA device enforcement

Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belg...

Official source for Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions:...
FDA device enforcement

Azurion 5 M20 System Model Numbers: (1)722228, (2)722232, (3)722281(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belg...

Official source for Azurion 5 M20 System Model Numbers: (1)722228, (2)722232, (3)722281(OUS only); Software Version N...
FDA device enforcement

Product Name: Allura Xper FD10C; Model number: 722001;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Product Name: Allura Xper FD10C; Model number: 722001;
FDA device enforcement

Allura Xper FD10F; Model Number: 722002;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD10F; Model Number: 722002;
FDA device enforcement

Allura Xper FD10; Model Number: 722003;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD10; Model Number: 722003;
FDA device enforcement

Allura Xper FD10/10; Model Number: 722005;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD10/10; Model Number: 722005;
FDA device enforcement

Allura Xper FD20; Model Number: 722006;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD20; Model Number: 722006;
FDA device enforcement

Allura Xper FD20 Biplane; Model Number: 722008;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD20 Biplane; Model Number: 722008;
FDA device enforcement

Allura Xper FD10; Model Number: 722010;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD10; Model Number: 722010;
FDA device enforcement

Allura Xper FD10/10; Model Number: 722011;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD10/10; Model Number: 722011;
FDA device enforcement

Allura Xper FD20; Model Number: 722012;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD20; Model Number: 722012;
FDA device enforcement

Allura Xper FD20 Biplane; Model Number: 722013;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD20 Biplane; Model Number: 722013;
FDA device enforcement

Allura Xper FD20 OR Table; Model Number: 722023;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD20 OR Table; Model Number: 722023;
FDA device enforcement

Allura Xper FD10; Model Number: 722026;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD10; Model Number: 722026;
FDA device enforcement

Allura Xper FD10/10; Model Number: 722027;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Re...

Official source for Allura Xper FD10/10; Model Number: 722027;

Reasons in these listed records

  • Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
  • Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for cool...

These examples come from this result page and are not a ranking of every recall reason.

Checking PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Is there a recall on philips medical systems nederland b.v. right now?

This archive contains 208 records, including 208 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking