FDA device enforcement

Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 12 mm 027 REF PED2-300-12. Intracranial aneurysm flow diverter

Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of CT, DE, IN, MA, NC, NJ, NM, NY, OH, PA, TX, VA, WA.

Official source for Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 12 mm 027 REF PED2-300-12. In...
FDA device enforcement

Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 20 mm 027 REF PED2-300-20. Intracranial aneurysm flow diverter

Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of CT, DE, IN, MA, NC, NJ, NM, NY, OH, PA, TX, VA, WA.

Official source for Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 20 mm 027 REF PED2-300-20. In...
FDA device enforcement

Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12,...

Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month im...

Class IReported ongoing

Distribution: Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, ...

Official source for Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3...
FDA device enforcement

Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20,...

Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month im...

Class IReported ongoing

Distribution: Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, ...

Official source for Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3...
FDA device enforcement

The Apollo Onyx Delivery Microcatheter (AOMC) is a DMSO-compatible, single-lumen, end hole catheter designed for the infusion of the Onyx liquid embolic system (LES). The cathet...

European version of microcatheter were distributed within US which contain a different "Indications for Use".

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, HI, IA, IL, IN, KS, MA, MD, MI, MO, NC, ND, NE, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI.

Official source for The Apollo Onyx Delivery Microcatheter (AOMC) is a DMSO-compatible, single-lumen, end hole cathet...
FDA device enforcement

The Apollo Onyx Delivery Microcatheter (AOMC) is a DMSO-compatible, single-lumen, end hole catheter designed for the infusion of the Onyx liquid embolic system (LES). The cathet...

European version of microcatheter were distributed within US which contain a different "Indications for Use".

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, HI, IA, IL, IN, KS, MA, MD, MI, MO, NC, ND, NE, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI.

Official source for The Apollo Onyx Delivery Microcatheter (AOMC) is a DMSO-compatible, single-lumen, end hole cathet...

Reasons in these listed records

  • Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.
  • Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, pot...
  • European version of microcatheter were distributed within US which contain a different "Indications for Use".
  • Outer carton packaging of microcatheters were exposed to wet substance during transport.

These examples come from this result page and are not a ranking of every recall reason.

Checking Micro Therapeutics, Inc.

Is there a recall on micro therapeutics, inc. right now?

This archive contains 9 records, including 9 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking