Maquet Medical Systems USA recall records

35 reported ongoing

35 records in this archive. Latest source date Apr 15, 2024.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

QUADROX-iR, used during cardiopulmonary bypass Model Item No. Description BEQ-HMO 51100-USA 701067936 Small Adult with filter, with BIOLINE Coating BEQ-HMO 71100-USA 701067880 A...

1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging...

Class IIReported ongoing

Distribution: Domestic distribution nationwide.

Official source for QUADROX-iR, used during cardiopulmonary bypass Model Item No. Description BEQ-HMO 51100-USA 70106...
FDA device enforcement

QUADROX-i Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 30000-USA 701070384 HMO 31000-USA 701070388

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Class IIReported ongoing

Distribution: Nationwide domestic distribution

Official source for QUADROX-i Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 30000-USA 7...
FDA device enforcement

QUADROX-iD Adult, Oxygenator used during cardiopulmonary bypass Model Item No. HMOD 70000-USA 701067840 BEQ-HMOD70000-USA 701067859

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Class IIReported ongoing

Distribution: Nationwide domestic distribution

Official source for QUADROX-iD Adult, Oxygenator used during cardiopulmonary bypass Model Item No. HMOD 70000-USA 701...
FDA device enforcement

QUADROX-i Small Adult, Oxygenator used during cardiopulmonary bypass Model Item no. BEQ-HMO 50000-USA 701067905 HMO 50000-USA 701067891 HMO 51000-USA 701067895

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Class IIReported ongoing

Distribution: Nationwide domestic distribution

Official source for QUADROX-i Small Adult, Oxygenator used during cardiopulmonary bypass Model Item no. BEQ-HMO 50000...
FDA device enforcement

QUADROX-iD Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. BEQ-HMOD30000-USA 701070397 BEQ-HMOD30000-USA 701050330

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Class IIReported ongoing

Distribution: Nationwide domestic distribution

Official source for QUADROX-iD Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. BEQ-HMOD30000-...
FDA device enforcement

QUADROX-i Adult, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 70000-USA 701067820 HMO 71000-USA 701067823 BEQ-HMO 71000-USA 701067829

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Class IIReported ongoing

Distribution: Nationwide domestic distribution

Official source for QUADROX-i Adult, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 70000-USA 70106...
FDA device enforcement

QUADROX -i Neonatal, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 10000-USA 701070412 HMO 11000-USA 701070416

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Class IIReported ongoing

Distribution: Nationwide domestic distribution

Official source for QUADROX -i Neonatal, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 10000-USA 7...
FDA device enforcement

Venous Hardshell Cardiotomy Reservoir, used during cardiopulmonary bypass Model Item No. VKMO 10000-USA 701070441 VKMO 11000-USA 701070445

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Class IIReported ongoing

Distribution: Nationwide domestic distribution

Official source for Venous Hardshell Cardiotomy Reservoir, used during cardiopulmonary bypass Model Item No. VKMO 100...
FDA device enforcement

QUADROX-iR, used during cardiopulmonary bypass Model Item No. Description BEQ-HMO 51100-USA 701067936 Small Adult with filter, with BIOLINE coating BEQ-HMO 71100-USA 701067880 A...

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Class IIReported ongoing

Distribution: Nationwide domestic distribution

Official source for QUADROX-iR, used during cardiopulmonary bypass Model Item No. Description BEQ-HMO 51100-USA 70106...
FDA device enforcement

BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code 70106.9077, BEQ-HLS 7050 USA; HLS Set Advanced 7.0, Product Code 70106.9078. Part of the CARDIOHELP System, a pre-connected ...

Firm has initiated a removal of the product due to insufficient evidence of packaging sterility. Exposure to a non-sterile medical device may result in inflammation, infection, sepsis, or ischemia.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Braz...

Official source for BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code 70106.9077, BEQ-HLS 7050 USA; HLS Set Advanc...
FDA device enforcement

BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Part No. 70105.2794. A preconnected set used for extracorporeal respiratory and/or cardiovascular support, as part of the CARDIOHELP System.

The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020).

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries Canada and Australia.

Official source for BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Part No. 70105.2794. A preconnected set used for extracorp...
FDA device enforcement

BEQ-HLS 5050 USA; HLS Set Advanced 5.0 Part No. 70105.2797. A preconnected set used for extracorporeal respiratory and/or cardiovascular support, as part of the CARDIOHELP System.

The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020).

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries Canada and Australia.

Official source for BEQ-HLS 5050 USA; HLS Set Advanced 5.0 Part No. 70105.2797. A preconnected set used for extracorp...
FDA device enforcement

BEQ-HLS 5050 USA; HLS Set Advanced 5.0 Part No. 70106.9077. A preconnected set used for extracorporeal respiratory and/or cardiovascular support, as part of the CARDIOHELP System.

The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020).

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries Canada and Australia.

Official source for BEQ-HLS 5050 USA; HLS Set Advanced 5.0 Part No. 70106.9077. A preconnected set used for extracorp...
FDA device enforcement

BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Part No. 70106.9078. A preconnected set used for extracorporeal respiratory and/or cardiovascular support, as part of the CARDIOHELP System.

The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020).

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries Canada and Australia.

Official source for BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Part No. 70106.9078. A preconnected set used for extracorp...
FDA device enforcement

Cardiohelp-i Transport Guard with Transport Guard (701072525): is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circ...

Potential for a loose cylindrical pin on the Transport Guard that may result in the jamming of the locking mechanism causing the Transport Guard to not open or, if the locking pin is fully disengaged from the assembly, a full disassembly of the Transport Gu...

Class IIReported ongoing

Distribution: US Nationwide Distribution: FL, KY, NH, NJ, NY, PA, TX, VA Foreign: Belgium, Canada, Germany, Hungary, Luxemburg, Malaysia, Mexico, Poland, Slovakia, South Africa, South Korea, ...

Official source for Cardiohelp-i Transport Guard with Transport Guard (701072525): is a blood oxygenation and carbon ...

Reasons in these listed records

  • 1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to no...
  • Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.
  • Firm has initiated a removal of the product due to insufficient evidence of packaging sterility. Exposure to a non-sterile medical device may result in inflammation, infection, sepsis, or ischemia.
  • The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020).

These examples come from this result page and are not a ranking of every recall reason.

Checking Maquet Medical Systems USA

Is there a recall on maquet medical systems usa right now?

This archive contains 35 records, including 35 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking