Instrumentation Laboratory recall records

43 reported ongoing

43 records in this archive. Latest source date May 21, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

HemosIL ReadiPlasTin, Part No. 0020301400

Variable and out of specification QC results with HemosIL ReadiPlasTin. IL will remove all lots from the market and convert customers to an alternative prothrombin time (PT) reagent, specifically HemosIL RecombiPlasTin 2G.

Class IIReported ongoing

Distribution: US Nationwide distribution including in the states of AZ, CA, IA, IL, IN, KS, MD, ME, MI, MN, MO, NC, NE, NY, OH, OK, SC, TX, VA, WV.

Official source for HemosIL ReadiPlasTin, Part No. 0020301400
FDA device enforcement

ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS 00000280060R (Refurbished) 3. ACL TOP 350 CTS 00000280065 4. AC...

Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid A...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Ric...

Official source for ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2....
FDA device enforcement

HemosIL ReadiPlasTin (20 mL), Part number 0020301400. An in vitro diagnostic thromboplastin reagent.

The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of AZ, CA, IA, IL, KS, ME, MI, NC, NY, OK, SC, and VA. The countries of Canada, Chile, China, Costa Rica, Italy...

Official source for HemosIL ReadiPlasTin (20 mL), Part number 0020301400. An in vitro diagnostic thromboplastin reagent.

Reasons in these listed records

  • Variable and out of specification QC results with HemosIL ReadiPlasTin. IL will remove all lots from the market and convert customers to an alternative prothrombin time (PT) reagent, specifically HemosIL RecombiPlasTin 2G.
  • Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and Hemo...
  • The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential...

These examples come from this result page and are not a ranking of every recall reason.

Checking Instrumentation Laboratory

Is there a recall on instrumentation laboratory right now?

This archive contains 43 records, including 43 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking