Hamilton Medical, Inc. recall records

10 reported ongoing

10 records in this archive. Latest source date Jul 26, 2023.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Hamilton-C1, REF: 161001, Portable Intensive Care Ventilator

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Class IReported ongoing

Distribution: Distribution US nationwide, including Puerto Rico. OUS: Not provided

Official source for Hamilton-C1, REF: 161001, Portable Intensive Care Ventilator
FDA device enforcement

Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Class IReported ongoing

Distribution: Distribution US nationwide, including Puerto Rico. OUS: Not provided

Official source for Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator
FDA device enforcement

Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Class IReported ongoing

Distribution: Distribution US nationwide, including Puerto Rico. OUS: Not provided

Official source for Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator
FDA device enforcement

Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Class IReported ongoing

Distribution: Distribution US nationwide, including Puerto Rico. OUS: Not provided

Official source for Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
FDA device enforcement

Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Class IReported ongoing

Distribution: Distribution US nationwide, including Puerto Rico. OUS: Not provided

Official source for Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator
FDA device enforcement

HAMILTON C1 Ventilator REF 161001 1610010 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates

Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.

Class IReported ongoing

Distribution: U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware Florida Georgia Hawaii Idaho Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massac...

Official source for HAMILTON C1 Ventilator REF 161001 1610010 Is intended to provide positive pressure ventilatory su...
FDA device enforcement

HAMILTON T1 Ventilator REF 161006 1610060 161009 1610090 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates

Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.

Class IReported ongoing

Distribution: U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware Florida Georgia Hawaii Idaho Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massac...

Official source for HAMILTON T1 Ventilator REF 161006 1610060 161009 1610090 Is intended to provide positive pressure...
FDA device enforcement

HAMILTON C2 Ventilator Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates

Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.

Class IReported ongoing

Distribution: U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware Florida Georgia Hawaii Idaho Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massac...

Official source for HAMILTON C2 Ventilator Is intended to provide positive pressure ventilatory support to adults and...
FDA device enforcement

HAMILTON C3 Ventilator REF 160005 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates

Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.

Class IReported ongoing

Distribution: U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware Florida Georgia Hawaii Idaho Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massac...

Official source for HAMILTON C3 Ventilator REF 160005 Is intended to provide positive pressure ventilatory support to...

Reasons in these listed records

  • Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
  • Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.

These examples come from this result page and are not a ranking of every recall reason.

Checking Hamilton Medical, Inc.

Is there a recall on hamilton medical, inc. right now?

This archive contains 10 records, including 10 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking