FDA device enforcement

Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402-01, (7) M14050-00 (8) M14001;

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the material...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Po...

Official source for Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3...
FDA device enforcement

Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M15001, (7) M15002; Product Description: Th...

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the material...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Po...

Official source for Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3...
FDA device enforcement

Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Catalog Numbers: (1) 1530101, (2) 1530102 Product Description: The mobile hoist is ...

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the material...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Po...

Official source for Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Cat...
FDA device enforcement

Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number: 1430116,1430117,1430115 Product Description: The mobile hoist is an...

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the material...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Po...

Official source for Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Cat...

Reasons in these listed records

  • A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between t...

These examples come from this result page and are not a ranking of every recall reason.

Checking ETAC A/S

Is there a recall on etac a/s right now?

This archive contains 4 records, including 4 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking