Diversatek Healthcare recall records

6 reported ongoing

7 records in this archive. Latest source date Mar 18, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Diversatek Healthcare Viper¿ 3-Stage Fixed Wire Balloon Dilator, 6mm-7mm-8mm, REF 1204-08

The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.

Class IIReported ongoing

Distribution: US: MO, CA, FL, TX, AZ, PA, RI, NC, IL, MS, KS, NY, OH, VA, LA, IA, MI, NJ, AL, Brazil, Belgium

Official source for Diversatek Healthcare Viper¿ 3-Stage Fixed Wire Balloon Dilator, 6mm-7mm-8mm, REF 1204-08
FDA device enforcement

Diversatek Healthcare Viper¿ 3-Stage Wire Guided Balloon Dilator, 12mm-13.5mm-15mm, REF 1205-15

The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.

Class IIReported ongoing

Distribution: US: MO, CA, FL, TX, AZ, PA, RI, NC, IL, MS, KS, NY, OH, VA, LA, IA, MI, NJ, AL, Brazil, Belgium

Official source for Diversatek Healthcare Viper¿ 3-Stage Wire Guided Balloon Dilator, 12mm-13.5mm-15mm, REF 1205-15
FDA device enforcement

Diversatek Healthcare Viper¿ 3-Stage Wire Guided Balloon Dilator, 15mm-16.5mm-18mm, REF 1205-18

The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.

Class IIReported ongoing

Distribution: US: MO, CA, FL, TX, AZ, PA, RI, NC, IL, MS, KS, NY, OH, VA, LA, IA, MI, NJ, AL, Brazil, Belgium

Official source for Diversatek Healthcare Viper¿ 3-Stage Wire Guided Balloon Dilator, 15mm-16.5mm-18mm, REF 1205-18
FDA device enforcement

Diversatek Healthcare Viper¿ 3-Stage Wire Guided Balloon Dilator, 18mm-19mm-20mm, REF 1205-20

The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.

Class IIReported ongoing

Distribution: US: MO, CA, FL, TX, AZ, PA, RI, NC, IL, MS, KS, NY, OH, VA, LA, IA, MI, NJ, AL, Brazil, Belgium

Official source for Diversatek Healthcare Viper¿ 3-Stage Wire Guided Balloon Dilator, 18mm-19mm-20mm, REF 1205-20
FDA device enforcement

Zvu Anorectal Manometry Procedure Kit, REF: HRAM-PROC-10

Anorectal manometry procedure kit outer kit label is mislabeled with an incorrect expiration date; the manufacturing date was used in place of the expiration date. All inner labels on kit components display the correct expiration date.

Class IIIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of NM, MA, CO, OH and the countries of Mongolia, South Korea.

Official source for Zvu Anorectal Manometry Procedure Kit, REF: HRAM-PROC-10
FDA device enforcement

Diversatek Health, M-Catch Retrieval Nets, Part Number REF 1175-01, Quantity 5 EA/u. Used to endoscopically retrieve severed polyps, tissue samples, food bolus, and foreign bodi...

M-Catch Retrieval Nets, Part Number 1175-01 / Lot Number 001919, that were shipped between 9-27-2021 and 10-22-2021, may contain incorrect inner polybag labeling. The outer package labeling is correct.

Class IIReported terminated

Distribution: US Nationwide distribution.

Official source for Diversatek Health, M-Catch Retrieval Nets, Part Number REF 1175-01, Quantity 5 EA/u. Used to endo...

Reasons in these listed records

  • The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.
  • Anorectal manometry procedure kit outer kit label is mislabeled with an incorrect expiration date; the manufacturing date was used in place of the expiration date. All inner labels on kit components display the correct expiration date.
  • GI Software labelled as revision 3.3.0 is incorrectly labeled and is in fact revision 3.2.0.
  • M-Catch Retrieval Nets, Part Number 1175-01 / Lot Number 001919, that were shipped between 9-27-2021 and 10-22-2021, may contain incorrect inner polybag labeling. The outer package labeling is correct.

These examples come from this result page and are not a ranking of every recall reason.

Checking Diversatek Healthcare

Is there a recall on diversatek healthcare right now?

This archive contains 7 records, including 6 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking