FDA device enforcement

Semi-Automated External Defibrillators; Models: DDU-100 and DDU-100E. Item numbers: DCF-E110SG-DE/1 (German language), DCF-E110SG-FR/1 (French language), DCF-E110SG-IT/1 (Italia...

It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switze...

Class IIReported ongoing

Distribution: International Only: Switzerland.

Official source for Semi-Automated External Defibrillators; Models: DDU-100 and DDU-100E. Item numbers: DCF-E110SG-DE...
FDA device enforcement

Automated External Defibrillators, Model DDU-2200, IItem numbers: DCF-E2210-D3/1¿(German/English dual language), DCF-E2210-F3/1¿(French/English dual language)

It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switze...

Class IIReported ongoing

Distribution: International Only: Switzerland.

Official source for Automated External Defibrillators, Model DDU-2200, IItem numbers: DCF-E2210-D3/1¿(German/English ...
FDA device enforcement

Automated External Defibrillators, Model DDU-2300. Item numbers: DCF-E2310D3/1¿(German/English dual language), DCF-E2310E3/1¿(German/French dual language), DCF-E2310F3/1¿ (Frenc...

It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switze...

Class IIReported ongoing

Distribution: International Only: Switzerland.

Official source for Automated External Defibrillators, Model DDU-2300. Item numbers: DCF-E2310D3/1¿(German/English du...
FDA device enforcement

Automated External Defibrillators, Model DDU-2450. Item number: DCF-E2460DE (German language)

It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switze...

Class IIReported ongoing

Distribution: International Only: Switzerland.

Official source for Automated External Defibrillators, Model DDU-2450. Item number: DCF-E2460DE (German language)
FDA device enforcement

Automated External Defibrillators, Model DDU-2200, IItem numbers: DCF-E2210-D3/1¿(German/English dual language), DCF-E2210-F3/1¿(French/English dual language)

It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switze...

Class IIReported ongoing

Distribution: OUS only: Switzerland.

Official source for Automated External Defibrillators, Model DDU-2200, IItem numbers: DCF-E2210-D3/1¿(German/English ...
FDA device enforcement

Defibtech DDP-100 Adult Defibrillation Pads For use with AED Models: Lifeline and Lifeline Auto (DDU-1:XX Series)- Indicated for use on victims of sudden cardiac arrest (SCA)

Within the Irish Market, a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could lead to expired Defibtech AED Pads appearin...

Class IIReported ongoing

Distribution: Ireland

Official source for Defibtech DDP-100 Adult Defibrillation Pads For use with AED Models: Lifeline and Lifeline Auto (...
FDA device enforcement

Defibtech DDP-200P: Pediatric Defibrillation Pads 0 For use with AED Models: Lifeline and Lifeline Auto (DDU-l XX Series)-Indicated for use on victims of sudden cardiac arrest (...

Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech ...

Class IIReported ongoing

Distribution: Ireland

Official source for Defibtech DDP-200P: Pediatric Defibrillation Pads 0 For use with AED Models: Lifeline and Lifelin...
FDA device enforcement

Defibtech DDP-2001: Adult Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, lifeline ECG, lifeline View (DDU-2XXX Series). -Indicated for use on vic...

Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech ...

Class IIReported ongoing

Distribution: Ireland

Official source for Defibtech DDP-2001: Adult Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeli...
FDA device enforcement

Defibtech DDP-2002: Pediatric Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, Lifeline ECG, Lifeline View (DDU-2XXX Series -Indicated for use on v...

Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech ...

Class IIReported ongoing

Distribution: Ireland

Official source for Defibtech DDP-2002: Pediatric Defibrillation Pads For use with AED Models: Lifeline View Auto, li...
FDA device enforcement

DDU-100 Lifeline Automated External Defibrillator (AED)-Indicated for use on victims of sudden cardiac arrest

A component of a sub-assembly used in the affected AEDs had not undergone durability and reliability testing, and if a malfunction occurs, therefore the device could fail to deliver therapy.

Class IIReported ongoing

Distribution: US Nationwide distribution.

Official source for DDU-100 Lifeline Automated External Defibrillator (AED)-Indicated for use on victims of sudden ca...

Reasons in these listed records

  • It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switzerland (German, Frenc...
  • Within the Irish Market, a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could lead to expired Defibtech AED Pads appearing as though they are...
  • Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech AED Pads appearing a...
  • Problem in the device's motor may cause the device to stop compressions. This may lead to a delay in therapy if the device fails to operate, and could lead to patient injury or death.

These examples come from this result page and are not a ranking of every recall reason.

Checking Defibtech, LLC

Is there a recall on defibtech, llc right now?

This archive contains 11 records, including 11 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking