Records and reported status

FDA device enforcement

1202 Flex Focus Ultrasound System, cart models UA1214 and UA1814, used with: Flex Focus 200, Flex Focus 300, Flex Focus 400, Flex Focus 500, Flex Focus 700, Flex Focus 800; Syst...

GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.

Class IIReported ongoing

Distribution: Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cos...

Official source for 1202 Flex Focus Ultrasound System, cart models UA1214 and UA1814, used with: Flex Focus 200, Flex...
FDA device enforcement

Ultrasound System 1300, bkSpecto w/battery; Model No. 1300-21 or 1300-S1; System, Imaging, Pulsed Doppler, Ultrasonic

GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.

Class IIReported ongoing

Distribution: Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cos...

Official source for Ultrasound System 1300, bkSpecto w/battery; Model No. 1300-21 or 1300-S1; System, Imaging, Pulsed...
FDA device enforcement

UUltrasound System 2300, bkActiv w/battery; Model No. 2300-11; System, Imaging, Pulsed Doppler, Ultrasonic

GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.

Class IIReported ongoing

Distribution: Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cos...

Official source for UUltrasound System 2300, bkActiv w/battery; Model No. 2300-11; System, Imaging, Pulsed Doppler, U...
FDA device enforcement

bk3000 Ultrasound System w/battery; Model No. 2300-66; System, Imaging, Pulsed Doppler, Ultrasonic

GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.

Class IIReported ongoing

Distribution: Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cos...

Official source for bk3000 Ultrasound System w/battery; Model No. 2300-66; System, Imaging, Pulsed Doppler, Ultrasonic
FDA device enforcement

bk3500 Ultrasound System w/battery; Model No. 2300-41; System, Imaging, Pulsed Doppler, Ultrasonic

GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.

Class IIReported ongoing

Distribution: Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cos...

Official source for bk3500 Ultrasound System w/battery; Model No. 2300-41; System, Imaging, Pulsed Doppler, Ultrasonic
FDA device enforcement

bk5000 Ultrasound System w/battery; Model No. 2300-61; System, Imaging, Pulsed Doppler, Ultrasonic

GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.

Class IIReported ongoing

Distribution: Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cos...

Official source for bk5000 Ultrasound System w/battery; Model No. 2300-61; System, Imaging, Pulsed Doppler, Ultrasonic
FDA device enforcement

bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66, ultrasound system with battery option.

The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide - There was also government/military distribution and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Costa Rica, Croati...

Official source for bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model ...

Reasons in these listed records

  • GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
  • The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.
  • An issue on the Dual Live Compare feature on the bkActiv system could result in a measurement error.

These examples come from this result page and are not a ranking of every recall reason.

Checking B-K Medical A/S

Is there a recall on b-k medical a/s right now?

This archive contains 8 records, including 8 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking