# REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments: - B1018-280 MicroScan... — FDA record Z-2889-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 26, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument
- **Reason**: Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
- **Classification**: Class II
- **Recalling firm**: Beckman Coulter Inc.
- **Quantity**: Not supplied
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-2889-2026

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2889-2026%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | UDI: 15099590799250/ V5.03 Panel Update 11 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WI, WV, WY.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Alaska](https://isitonrecall.com/state/alaska/)
- [Alabama](https://isitonrecall.com/state/alabama/)
- [Arkansas](https://isitonrecall.com/state/arkansas/)
- [Arizona](https://isitonrecall.com/state/arizona/)
- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Connecticut](https://isitonrecall.com/state/connecticut/)
- [Delaware](https://isitonrecall.com/state/delaware/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Georgia](https://isitonrecall.com/state/georgia/)
- [Hawaii](https://isitonrecall.com/state/hawaii/)
- [Iowa](https://isitonrecall.com/state/iowa/)
- [Idaho](https://isitonrecall.com/state/idaho/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Indiana](https://isitonrecall.com/state/indiana/)
- [Kansas](https://isitonrecall.com/state/kansas/)
- [Kentucky](https://isitonrecall.com/state/kentucky/)
- [Louisiana](https://isitonrecall.com/state/louisiana/)
- [Massachusetts](https://isitonrecall.com/state/massachusetts/)
- [Maryland](https://isitonrecall.com/state/maryland/)
- [Maine](https://isitonrecall.com/state/maine/)
- [Michigan](https://isitonrecall.com/state/michigan/)
- [Minnesota](https://isitonrecall.com/state/minnesota/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [Mississippi](https://isitonrecall.com/state/mississippi/)
- [Montana](https://isitonrecall.com/state/montana/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [North Dakota](https://isitonrecall.com/state/north-dakota/)
- [Nebraska](https://isitonrecall.com/state/nebraska/)
- [New Hampshire](https://isitonrecall.com/state/new-hampshire/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [New Mexico](https://isitonrecall.com/state/new-mexico/)
- [Nevada](https://isitonrecall.com/state/nevada/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Ohio](https://isitonrecall.com/state/ohio/)
- [Oklahoma](https://isitonrecall.com/state/oklahoma/)
- [Oregon](https://isitonrecall.com/state/oregon/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [Rhode Island](https://isitonrecall.com/state/rhode-island/)
- [South Carolina](https://isitonrecall.com/state/south-carolina/)
- [South Dakota](https://isitonrecall.com/state/south-dakota/)
- [Tennessee](https://isitonrecall.com/state/tennessee/)
- [Texas](https://isitonrecall.com/state/texas/)
- [Utah](https://isitonrecall.com/state/utah/)
- [Virginia](https://isitonrecall.com/state/virginia/)
- [Vermont](https://isitonrecall.com/state/vermont/)
- [Washington](https://isitonrecall.com/state/washington/)
- [Wisconsin](https://isitonrecall.com/state/wisconsin/)
- [West Virginia](https://isitonrecall.com/state/west-virginia/)
- [Wyoming](https://isitonrecall.com/state/wyoming/)

## Dates in the source

- **Report date**: Aug 12, 2026
- **Recall initiation date**: Jun 26, 2026
- **Center classification date**: Aug 5, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2889-2026%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Beckman Coulter Inc. records](https://isitonrecall.com/brand/beckman-coulter-inc/)