title: "Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series. Labele... — FDA record Z-2504-2025 | Is It on Recall"
description: "Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series."
url: https://isitonrecall.com/recall/fda-device-z-2504-2025-interventional-fluoroscopic-x-ray-system-philips-allura-xper-series-labeled-as-the-following-with-co/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series. Labeled as the following with corresponding mo... — FDA record Z-2504-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 1, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series. Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003. 2. Allura Xper FD10/10, Model Number 722005. 3. Allura Xper FD20, Model Number 722006. 4. Allura Xper FD20 Biplane, Model Number 722008. 5. Allura Xper FD10, Model Number 722010. 6. Allura Xper FD10/10, Model Number 722011. 7. Allura Xper FD20, Model Number 722012. 8. Allura Xper FD20 Biplane, Model Number 722013. 9. Allura Xper FD20 OR Table, Model Number 722023. 10. Allura Xper FD10, Model Number 722026. 11. Allura Xper FD10/10, Model Number 722027. 12. Allura Xper FD20, Model Number 722028. 13. Allura Xper FD20/10, Model Number 722029. 14. Allura Xper FD20 OR Table, Model Number 722035. 15. Allura Xper FD20/20, Model Number 722038. 16. Allura Xper FD20/15, Model Number 722058. 510(k) Numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859.
- **Reason**: Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
- **Classification**: Class II
- **Recalling firm**: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- **Quantity**: 192 untis
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-2504-2025

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2504-2025%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | 1. Model Number/System Code (UDI); Serial Numbers: 722003 (No UDI); 292, 314, 1077, 464. 2. Model Number/System Code (UDI); Serial Numbers: 722005 (No UDI); 54. 3. Model Number/System Code (UDI); Serial Numbers: 722006 (No UDI); 2126, 2200, 1560, 1921, 1724, 1219. 4. Model Number/System Code (UDI); Serial Numbers: 722008 (No UDI); 516, 142, 434, 418, 480. 5. Model Number/System Code (UDI); Serial Numbers: 722010 (00884838059030); 33, 968, 763, 215, 1425, 1126, 818, 1028, 1043, 608, 907, 50, 1390, 706, 900, 1439, 1152, 165, 62, 1031, 1088, 741, 840, 413. 6. Model Number/System Code (UDI); Serial Numbers: 722011 (00884838059047); 308, 344, 243, 266, 231, 286, 369, 378, 259, 333. 7. Model Number/System Code (UDI); Serial Numbers: 722012 (00884838059054); 2898, 2896, 1594, 1249, 2057, 2135, 2857, 792, 2841, 1334, 1550, 2868, 2912, 1094, 2282, 1681, 1332, 2032, 2076, 176, 835, 626, 1320, 1439, 1667. 8. Model Number/System Code (UDI); Serial Numbers: 722013 (00884838059061); 115, 575, 272, 632, 378, 608, 412, 353, 595, 355, 403. 9. Model Number/System Code (UDI); Serial Numbers: 722023 (00884838059085); 128, 83, 86, 134, 87. 10. Model Number/System Code (UDI); Serial Numbers: 722026 (00884838054189); 233, 561, 942, 368, 826, 1115, 539, 159, 598, 365, 528, 603, 263, 792, 783, 679, 680, 406, 388, 302, 659, 736, 1022, 22, 506, 120, 526. 11. Model Number/System Code (UDI); Serial Numbers: 722027 (00884838054196); 106, 127, 78, 204, 113, 82, 66, 119. 12. Model Number/System Code (UDI); Serial Numbers: 722028 (00884838054202); 668, 265, 1154, 180, 880, 274, 172, 1111, 806, 451, 236, 887, 874, 1266, 1499, 1514, 1457, 908, 1292, 247, 944, 951, 1390, 1662, 415, 105, 1675, 348, 1562, 1029, 51, 941, 403, 427, 659, 935, 1225, 1165, 1621, 725, 723, 1559, 753, 883. 13. Model Number/System Code (UDI); Serial Numbers: 722029 (00884838054219); 69, 58. 14. Model Number/System Code (UDI); Serial Numbers: 722035 (00884838054240); 187, 125, 40, 108, 163, 26, 75, 173, 190, 103. 15. Model Number/System Code (UDI); Serial Numbers: 722038 (00884838054226); 63. 16. Model Number/System Code (UDI); Serial Numbers: 722058 (00884838059115); 238, 161, 84, 39, 133, 244, 193, 187, 102. |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Worldwide - US Nationwide distribution in the states of AL, AZ, CA, DC, FL, GA, HI, ID, IL, KS, KY, LA, MI, MN, MO, NC, NJ, NM, NV, NY, OH, OR, PA, TN, TX, UT, VA, WI, WV and the countries of Argentina, Austria, Australia, Belgium, Switzerland, Chile, China, Colombia, Czechia, Germany, Egypt, Spain, Finland, France, United Kingdom, Hong Kong, India, Italy, Japan, Korea, Republic of, Malaysia, Netherlands, Norway, Oman, Pakistan, Poland, Romania, Saudi Arabia, Sweden, Singapore, Thailand, Vietnam.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Alaska](https://isitonrecall.com/state/alaska/)
- [Alabama](https://isitonrecall.com/state/alabama/)
- [Arkansas](https://isitonrecall.com/state/arkansas/)
- [Arizona](https://isitonrecall.com/state/arizona/)
- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Connecticut](https://isitonrecall.com/state/connecticut/)
- [Delaware](https://isitonrecall.com/state/delaware/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Georgia](https://isitonrecall.com/state/georgia/)
- [Hawaii](https://isitonrecall.com/state/hawaii/)
- [Iowa](https://isitonrecall.com/state/iowa/)
- [Idaho](https://isitonrecall.com/state/idaho/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Indiana](https://isitonrecall.com/state/indiana/)
- [Kansas](https://isitonrecall.com/state/kansas/)
- [Kentucky](https://isitonrecall.com/state/kentucky/)
- [Louisiana](https://isitonrecall.com/state/louisiana/)
- [Massachusetts](https://isitonrecall.com/state/massachusetts/)
- [Maryland](https://isitonrecall.com/state/maryland/)
- [Maine](https://isitonrecall.com/state/maine/)
- [Michigan](https://isitonrecall.com/state/michigan/)
- [Minnesota](https://isitonrecall.com/state/minnesota/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [Mississippi](https://isitonrecall.com/state/mississippi/)
- [Montana](https://isitonrecall.com/state/montana/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [North Dakota](https://isitonrecall.com/state/north-dakota/)
- [Nebraska](https://isitonrecall.com/state/nebraska/)
- [New Hampshire](https://isitonrecall.com/state/new-hampshire/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [New Mexico](https://isitonrecall.com/state/new-mexico/)
- [Nevada](https://isitonrecall.com/state/nevada/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Ohio](https://isitonrecall.com/state/ohio/)
- [Oklahoma](https://isitonrecall.com/state/oklahoma/)
- [Oregon](https://isitonrecall.com/state/oregon/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [Rhode Island](https://isitonrecall.com/state/rhode-island/)
- [South Carolina](https://isitonrecall.com/state/south-carolina/)
- [South Dakota](https://isitonrecall.com/state/south-dakota/)
- [Tennessee](https://isitonrecall.com/state/tennessee/)
- [Texas](https://isitonrecall.com/state/texas/)
- [Utah](https://isitonrecall.com/state/utah/)
- [Virginia](https://isitonrecall.com/state/virginia/)
- [Vermont](https://isitonrecall.com/state/vermont/)
- [Washington](https://isitonrecall.com/state/washington/)
- [Wisconsin](https://isitonrecall.com/state/wisconsin/)
- [West Virginia](https://isitonrecall.com/state/west-virginia/)
- [Wyoming](https://isitonrecall.com/state/wyoming/)

## Dates in the source

- **Report date**: Sep 10, 2025
- **Recall initiation date**: Aug 1, 2025
- **Center classification date**: Sep 3, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2504-2025%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. records](https://isitonrecall.com/brand/philips-medical-systems-nederland-b-v/)