title: "Zimmer Periarticular Locking Plate (ZPLP) Distal Lateral Femoral Plate 6 hole... — FDA record Z-2254-2023 | Is It on Recall"
description: "Thread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recogniz"
url: https://isitonrecall.com/recall/fda-device-z-2254-2023-zimmer-periarticular-locking-plate-zplp-distal-lateral-femoral-plate-6-holes-159-mm-length-indicated/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Zimmer Periarticular Locking Plate (ZPLP) Distal Lateral Femoral Plate 6 holes 159 mm Length-Indicated for temporary ... — FDA record Z-2254-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 27, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Zimmer Periarticular Locking Plate (ZPLP) Distal Lateral Femoral Plate 6 holes 159 mm Length-Indicated for temporary internal fixation and stabilization of osteotomies and fractures, including comminuted fractures, supracondylar fractures, intra-articular and extra-articular condylar fractures, fractures in osteopenic bone, nonunions, and malunions. Item Number: 00235710106
- **Reason**: Thread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock.
- **Classification**: Class II
- **Recalling firm**: Zimmer, Inc.
- **Quantity**: 75 units
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-2254-2023

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2254-2023%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Lot Number/UDI: 64254511 /(01)0089024055940(10)64254511; 64296571/(01)0089024055940(10)64296571; 65379087/(01)0089024055940(10)65379087; 65379093/(01)0089024055940(10)65379093; 65379096/(01)0089024055940(10)65379096 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

AR AZ MA ME MO NJ NM NY OR WA Foreign: Brazil China Dominican Republic Korea Netherlamds Taiwan

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Arkansas](https://isitonrecall.com/state/arkansas/)
- [Arizona](https://isitonrecall.com/state/arizona/)
- [Massachusetts](https://isitonrecall.com/state/massachusetts/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [New Mexico](https://isitonrecall.com/state/new-mexico/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Washington](https://isitonrecall.com/state/washington/)

## Dates in the source

- **Report date**: Aug 9, 2023
- **Recall initiation date**: Jun 27, 2023
- **Center classification date**: Jul 28, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2254-2023%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Zimmer, Inc. records](https://isitonrecall.com/brand/zimmer-inc/)