title: "OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7... — FDA record Z-2213-2026 | Is It on Recall"
description: "Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant"
url: https://isitonrecall.com/recall/fda-device-z-2213-2026-onelif-interbody-cage-ref-description-010-107-1007-small-7-degree-10mm-x-7mm-010-107-1007-2-small-7/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X... — FDA record Z-2213-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 23, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-107-1209/Small 7 Degree 12mm X 9mm, 010-107-1411/Small 7 Degree 14mm X 11mm, 010-107-1613/Small 7 Degree 16mm X 13mm, 010-113-1006/Small, 13 Degree 10mm X 6mm, 010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique, 010-113-1208/Small 13 Degree 12mm X 8mm, 010-113-1410/Small 13 Degree 14mm X 10mm, 010-113-1612/Small 13 Degree 16mm X 12mm, 010-119-1407/19 Degree 14mm X 7mm, 010-119-1609/Small 19 Degree 16mm X 9mm, 010-119-1811/Small 19 Degree 18mm X 11mm, 010-125-1606/Small 25 Degree 16mm X 6mm, 010-125-1808/Small 25 Degree 18mm X 8mm, 010-207-1007/Medium 7 Degree 10mm X 7mm, 010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique, 010-207-1209/Medium 7 Degree 12mm X 9mm, 010-207-1411/Medium 7 Degree 14mm X 11mm, 010-207-1613/Medium 7 Degree 16mm X 13mm, 010-213-1206/Medium 13 Degree 12mm X 6mm, 010-213-1408/Medium 13 Degree 14mm X 8mm, 010-213-1610/Medium 13 Degree 16mm X 10mm, 010-213-1812/Medium 13 Degree 18mm X 12mm, 010-219-1406/Medium 19 Degree 14mm X 6mm, 010-219-1608/Medium 19 Degree 16mm X 8mm, 010-219-1810/Medium 19 Degree 18mm X 10mm, 010-225-1807/Medium 25 Degree 18mm X 7mm, 010-225-2009/Medium 25 Degree 20mm X 9mm, 010-307-1006/Large 7 Degree,10mm X 6mm, 010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique, 010-307-1208/ Large 7 Degree 12mm X 8mm, 010-307-1410/Large 7 Degree 14mm X 10mm, 010-307-1612/Large 7 Degree 16mm X 12mm, 010-313-1205/Large 13 Degree 12mm X 5mm, 010-313-1407/Large 13 Degree 14mm X 7mm, 010-313-1609/Large 13 Degree 16mm X 9mm, 010-313-1811/Large 13 Degree 18mm X 11mm, 010-313-2013/Large 13 Degree 20mm X 13mm, 010-319-1606/Large 19 Degree 16mm X 6mm, 010-319-1808/ Large 19 Degree 18mm X 8mm, 010-319-2010/ Large 19 Degree 20mm X 10mm, 010-325-2007/Large 25 Degree 20mm X 7mm, 010-325-2209/ Large 25 Degree 22mm X 9mm, Implant Kits containing OneLIF Interbody Cages, Name/Kit Family: Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm, Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLI...
- **Reason**: Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.
- **Classification**: Class II
- **Recalling firm**: Novapproach Spine, LLC
- **Quantity**: 1259
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-2213-2026

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2213-2026%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | REF/UDI-DI/ Cage Lots: 010-107-1007/G07001010710070/ 103003, 105701, 118001, 129301; 010-107-1007-2/G070010107100720/ 114301, 117509, 128703; 010-107-1209/G07001010712090/ 102714, 105702, 117501; 010-107-1411/G07001010714110/ 103019, 105703; 010-107-1613/G07001010716130/ 103020, 105704, 127701; 010-113-1006/G07001011310060/ 103021, 105705, 118002, 129801; 010-113-1006-2/G070010113100620/ 117510, 127716, 114302; 010-113-1208/G07001011312080/ 103022, 105706, 117502, 126902; 010-113-1410/G07001011314100/ 102715, 105707, 127702; 010-113-1612/G07001011316120/ 103023, 105708, 127703; 010-119-1407/G07001011914070/ 103024, 105709, 125101, 128701; 010-119-1609/G07001011916090/ 102716, 105710, 127001; 010-119-1811/G07001011918110/ 103025, 105711, 127003; 010-125-1606/G07001012516060/ 102717, 117901, 127002; 010-125-1808/G07001012518080/ 103005, 117902, 127704; 010-207-1007/G07001020710070/ 102702, 106201, 118003, 125102; 010-207-1007-2/G070010207100720/ 117511, 124106, 127717, 114303; 010-207-1209/G07001020712090/ 102701, 105401, 124101, 127705, 101201; 010-207-1411/G07001020714110/ 105001, 105402, 101203; 010-207-1613/G07001020716130/ 103007, 105403, 124102, 127706; 010-213-1206/G07001021312060/ 105002, 105404, 121202, 124103, 127707, 101204; 010-213-1408/G07001021314080/ 102704, 105405, 121203, 124104, 127708; 010-213-1610/G07001021316100/ 103008, 105406, 130101; 010-213-1812/G07001021318120/ 103009, 105407, 127709; 010-219-1406/G07001021914060/ 102705, 105408, 117503, 119301, 121201, 124105; 010-219-1608/G07001021916080/ 103010, 105409, 121204, 127710; 010-219-1810/G07001021918100/ 103011, 105410, 127711; 010-225-1807/G07001022518070/ 102706, 113701, 117001, 121205; 010-225-2009/G07001022520090/ 103001, 117903, 127004, 127712; 010-307-1006/G07001030710060/ 102707, 106202, 118004, 129302; 010-307-1006-2/G070010307100620/ 117512, 127718, 114304; 010-307-1208/G07001030712080/ 102709, 105601, 117504; 010-307-1410/G07001030714100/ 102710, 105602, 127713; 010-307-1612/G07001030716120/ 103012, 105603, 117505; 010-313-1205/G07001031312050/ 103013, 105604, 117506, 129303; 010-313-1407/G07001031314070/ 102711, 105605, 117507, 124107, 128702, 130102; 010-313-1609/G07001031316090/ 103014, 105606; 010-313-1811/G07001031318110/ 103015, 105607, 127714; 010-313-2013/G07001031320130/ 103016, 105608, 127715; 010-319-1606/G07001031916060/ 102712, 105609, 114905, 117508; 010-319-1808/G07001031918080/ 103017, 105610, 125103; 010-319-2010/G07001031920100/ 103018, 105611, 125104; 010-325-2007/G07001032520070/ 102713, 105612, 121206; 010-325-2209/G07001032522090/ 105613, 121207, 101206, 101207, 101208 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

US Nationwide distribution in the states of NY, CT, CA, TX, FL, MI.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Alaska](https://isitonrecall.com/state/alaska/)
- [Alabama](https://isitonrecall.com/state/alabama/)
- [Arkansas](https://isitonrecall.com/state/arkansas/)
- [Arizona](https://isitonrecall.com/state/arizona/)
- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Connecticut](https://isitonrecall.com/state/connecticut/)
- [Delaware](https://isitonrecall.com/state/delaware/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Georgia](https://isitonrecall.com/state/georgia/)
- [Hawaii](https://isitonrecall.com/state/hawaii/)
- [Iowa](https://isitonrecall.com/state/iowa/)
- [Idaho](https://isitonrecall.com/state/idaho/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Indiana](https://isitonrecall.com/state/indiana/)
- [Kansas](https://isitonrecall.com/state/kansas/)
- [Kentucky](https://isitonrecall.com/state/kentucky/)
- [Louisiana](https://isitonrecall.com/state/louisiana/)
- [Massachusetts](https://isitonrecall.com/state/massachusetts/)
- [Maryland](https://isitonrecall.com/state/maryland/)
- [Maine](https://isitonrecall.com/state/maine/)
- [Michigan](https://isitonrecall.com/state/michigan/)
- [Minnesota](https://isitonrecall.com/state/minnesota/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [Mississippi](https://isitonrecall.com/state/mississippi/)
- [Montana](https://isitonrecall.com/state/montana/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [North Dakota](https://isitonrecall.com/state/north-dakota/)
- [Nebraska](https://isitonrecall.com/state/nebraska/)
- [New Hampshire](https://isitonrecall.com/state/new-hampshire/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [New Mexico](https://isitonrecall.com/state/new-mexico/)
- [Nevada](https://isitonrecall.com/state/nevada/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Ohio](https://isitonrecall.com/state/ohio/)
- [Oklahoma](https://isitonrecall.com/state/oklahoma/)
- [Oregon](https://isitonrecall.com/state/oregon/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [Rhode Island](https://isitonrecall.com/state/rhode-island/)
- [South Carolina](https://isitonrecall.com/state/south-carolina/)
- [South Dakota](https://isitonrecall.com/state/south-dakota/)
- [Tennessee](https://isitonrecall.com/state/tennessee/)
- [Texas](https://isitonrecall.com/state/texas/)
- [Utah](https://isitonrecall.com/state/utah/)
- [Virginia](https://isitonrecall.com/state/virginia/)
- [Vermont](https://isitonrecall.com/state/vermont/)
- [Washington](https://isitonrecall.com/state/washington/)
- [Wisconsin](https://isitonrecall.com/state/wisconsin/)
- [West Virginia](https://isitonrecall.com/state/west-virginia/)
- [Wyoming](https://isitonrecall.com/state/wyoming/)

## Dates in the source

- **Report date**: May 27, 2026
- **Recall initiation date**: Apr 23, 2026
- **Center classification date**: May 20, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2213-2026%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Novapproach Spine LLC records](https://isitonrecall.com/brand/novapproach-spine-llc/)