title: "VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000 — FDA record Z-1942-2026 | Is It on Recall"
description: "Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions durin"
url: https://isitonrecall.com/recall/fda-device-z-1942-2026-vocsn-v-pro-package-ref-prt-01198-000-vocsn-v-pro-unit-ref-prt-01185-000/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000 — FDA record Z-1942-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 25, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
- **Reason**: Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk
- **Classification**: Class I
- **Recalling firm**: Ventec Life Systems, Inc.
- **Quantity**: 42
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-1942-2026

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1942-2026%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | UDI-DI (REF/UDI-DI): PRT-01198-000/00855573007914, PRT-01185-000/00855573007877 Package/Device Serial Number (package serial number may include a "P" at the end): 152884, 152885, 152886, 152887, 152888, 152889, 152890, 152891, 152892, 152893, 152895, 152896, 152897, 152898, 152899, 152900, 152901, 152902, 152903, 152904, 152908, 152909, 152910, 152911, 152912, 152913, 152914, 152915, 152916, 152917, 152918, 152919, 152986, 152987, 155236, 156507, 156512, 156515, 156516, 156517, 156524, 156532 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

US: OH, GA, CA, MO

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [California](https://isitonrecall.com/state/california/)
- [Georgia](https://isitonrecall.com/state/georgia/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [Ohio](https://isitonrecall.com/state/ohio/)

## Dates in the source

- **Report date**: May 13, 2026
- **Recall initiation date**: Mar 25, 2026
- **Center classification date**: May 1, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1942-2026%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Ventec Life Systems, Inc. records](https://isitonrecall.com/brand/ventec-life-systems-inc/)