# Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082085 — FDA record Z-1891-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 25, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082085
- **Reason**: Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- **Classification**: Class I
- **Recalling firm**: TELEFLEX LLC
- **Quantity**: 39550 units
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-1891-2023

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1891-2023%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | UDI/DI 14026704341457, Batch Numbers: 18HG37, 18HT15, 18HT24, 18JG12, 18JG31, 19CT69, 19CT73, 19CT74, 19CT76, 19CT77, 19DT05, 19DT10, 19KT14, 19KT30, 19KT46, 19KT49, 19KT50, 19KT55, 19LT35, 20BT51, 20CT02, 20CT22, 20CT28, 20CT32, 20CT42, 20DT02, 20DT17, 20DT18, 20ET17, 20GG43, KME20H0750, KME20J2539, KME20J2680, KME20K0923, KME20K1114, KME20K1513, KME20K3034, KME20K3146, KME20L1867, KME20L2274, KME20M1511, KME21E0902, KME21F0741, KME21F1214, KME21F2006, KME21F2074, KME21K0095, KME21K1193, KME21K2275, KME21K2575, KME21K3015, KME21L0394, KME21L0799, KME21L1115, KME21L1276, KME21L1586, KME21M2500, KME21M2728, KME22A0787, KME22A2287, KME22A2957, KME22A2968, KME22B0519, KME22B1154, KME22B1696, KME22B1775, KME22B2179, KME22B2320, KME22B2477, KME22B2797, KME22C2074, KME22D1891, KME22D1892, KME22D3186, KME22D3488, KME22D3564, KME22E0663, KME22E0988, KME22E1279, KME22E1583, KME22E1811, KME22E2253 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

US Nationwide distribution including Puerto Rico.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Alaska](https://isitonrecall.com/state/alaska/)
- [Alabama](https://isitonrecall.com/state/alabama/)
- [Arkansas](https://isitonrecall.com/state/arkansas/)
- [Arizona](https://isitonrecall.com/state/arizona/)
- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Connecticut](https://isitonrecall.com/state/connecticut/)
- [Delaware](https://isitonrecall.com/state/delaware/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Georgia](https://isitonrecall.com/state/georgia/)
- [Hawaii](https://isitonrecall.com/state/hawaii/)
- [Iowa](https://isitonrecall.com/state/iowa/)
- [Idaho](https://isitonrecall.com/state/idaho/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Indiana](https://isitonrecall.com/state/indiana/)
- [Kansas](https://isitonrecall.com/state/kansas/)
- [Kentucky](https://isitonrecall.com/state/kentucky/)
- [Louisiana](https://isitonrecall.com/state/louisiana/)
- [Massachusetts](https://isitonrecall.com/state/massachusetts/)
- [Maryland](https://isitonrecall.com/state/maryland/)
- [Maine](https://isitonrecall.com/state/maine/)
- [Michigan](https://isitonrecall.com/state/michigan/)
- [Minnesota](https://isitonrecall.com/state/minnesota/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [Mississippi](https://isitonrecall.com/state/mississippi/)
- [Montana](https://isitonrecall.com/state/montana/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [North Dakota](https://isitonrecall.com/state/north-dakota/)
- [Nebraska](https://isitonrecall.com/state/nebraska/)
- [New Hampshire](https://isitonrecall.com/state/new-hampshire/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [New Mexico](https://isitonrecall.com/state/new-mexico/)
- [Nevada](https://isitonrecall.com/state/nevada/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Ohio](https://isitonrecall.com/state/ohio/)
- [Oklahoma](https://isitonrecall.com/state/oklahoma/)
- [Oregon](https://isitonrecall.com/state/oregon/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [Rhode Island](https://isitonrecall.com/state/rhode-island/)
- [South Carolina](https://isitonrecall.com/state/south-carolina/)
- [South Dakota](https://isitonrecall.com/state/south-dakota/)
- [Tennessee](https://isitonrecall.com/state/tennessee/)
- [Texas](https://isitonrecall.com/state/texas/)
- [Utah](https://isitonrecall.com/state/utah/)
- [Virginia](https://isitonrecall.com/state/virginia/)
- [Vermont](https://isitonrecall.com/state/vermont/)
- [Washington](https://isitonrecall.com/state/washington/)
- [Wisconsin](https://isitonrecall.com/state/wisconsin/)
- [West Virginia](https://isitonrecall.com/state/west-virginia/)
- [Wyoming](https://isitonrecall.com/state/wyoming/)

## Dates in the source

- **Report date**: Jul 5, 2023
- **Recall initiation date**: May 25, 2023
- **Center classification date**: Jun 29, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1891-2023%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [TELEFLEX LLC records](https://isitonrecall.com/brand/teleflex-llc/)