# IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. In... — FDA record Z-1869-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 12, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.
- **Reason**: Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.
- **Classification**: Class I
- **Recalling firm**: Fresenius Kabi USA, LLC
- **Quantity**: 14,280 units (724 cases)
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-1869-2025

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1869-2025%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Model No. SET-0014-20; UDI 20811505030034; Lot No. FA24K05015. |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

US distribution: CA, CO, DE, GA, IL, MD, MI, MN, MS, NC, NJ, NV, OK, SC, TX, VA, WA & WI.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Delaware](https://isitonrecall.com/state/delaware/)
- [Georgia](https://isitonrecall.com/state/georgia/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Maryland](https://isitonrecall.com/state/maryland/)
- [Michigan](https://isitonrecall.com/state/michigan/)
- [Minnesota](https://isitonrecall.com/state/minnesota/)
- [Mississippi](https://isitonrecall.com/state/mississippi/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [Nevada](https://isitonrecall.com/state/nevada/)
- [Oklahoma](https://isitonrecall.com/state/oklahoma/)
- [South Carolina](https://isitonrecall.com/state/south-carolina/)
- [Texas](https://isitonrecall.com/state/texas/)
- [Virginia](https://isitonrecall.com/state/virginia/)
- [Washington](https://isitonrecall.com/state/washington/)
- [Wisconsin](https://isitonrecall.com/state/wisconsin/)

## Dates in the source

- **Report date**: Jun 18, 2025
- **Recall initiation date**: May 12, 2025
- **Center classification date**: Jun 6, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1869-2025%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Fresenius Kabi USA, LLC records](https://isitonrecall.com/brand/fresenius-kabi-usa-llc/)