# ARTIS Pheno VE30A and VE40A, Model 10849000 — FDA record Z-1814-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 12, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: ARTIS Pheno VE30A and VE40A, Model 10849000
- **Reason**: During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
- **Classification**: Class II
- **Recalling firm**: Siemens Medical Solutions USA, Inc
- **Quantity**: 73
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-1814-2026

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1814-2026%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | UDI 04056869046877 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

U.S. and O.U.S.

No US distribution state could be established from this record. Check the official notice.

## Dates in the source

- **Report date**: Apr 22, 2026
- **Recall initiation date**: Mar 12, 2026
- **Center classification date**: Apr 16, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.