---
title: "StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/... — FDA record Z-1666-2023 | Is It on Recall"
description: "In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occ"
url: https://isitonrecall.com/recall/fda-device-z-1666-2023-stealthstation-cranial-software-model-9735585-used-with-stealthstation-s7-i7-systems/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems — FDA record Z-1666-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 11, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems
- **Reason**: In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury
- **Classification**: Class I
- **Recalling firm**: Medtronic Navigation, Inc.
- **Quantity**: 556
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-1666-2023

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1666-2023%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Model/UDI-DI/Software Versions: 9735585/00763000253165, 00763000306762 (inside Kits: 00763000299927, 00763000299972), 00763000517267(inside Kits: 00763000517274, 00763000517335)/v3.1.1, 3.1.2, and 3.1.3 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

US Distribution: PR. OUS (foreign) distribution: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, Colombia COLOMBIA, C¿TE D'IVOIRE, CROATIA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KENYA, LATVIA, MEXICO, MOLDAVIA, NETHERLANDS, PAKISTAN, POLAND, PORTUGAL, REPUBLIC of KOREA, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED ARAB EMIRATES

No US distribution state could be established from this record. Check the official notice.

## Dates in the source

- **Report date**: Jun 14, 2023
- **Recall initiation date**: Apr 11, 2023
- **Center classification date**: Jun 2, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1666-2023%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Medtronic Navigation, Inc. records](https://isitonrecall.com/brand/medtronic-navigation-inc/)
