title: "Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for us... — FDA record Z-1663-2023 | Is It on Recall"
description: "Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts t"
url: https://isitonrecall.com/recall/fda-device-z-1663-2023-teleflex-horizon-microclip-titanium-ligating-clips-w-tape-ref-005200-for-use-in-general-surgical-pro/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that re... — FDA record Z-1663-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 28, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or tissue ligation
- **Reason**: Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures.
- **Classification**: Class II
- **Recalling firm**: TELEFLEX LLC
- **Quantity**: 146,606 units
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-1663-2023

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1663-2023%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | UDI/DI 24026704696394, Lot Numbers: 73A2200857, 73A2300018, 73A2300176, 73A2300326, 73A2300520, 73A2300778, 73D2200216, 73D2200559, 73E2200634, 73E2200874, 73E2201099, 73F2200193, 73F2200200, 73F2200482, 73F2200711, 73G2200168, 73G2200345, 73G2200560, 73G2200754, 73H2200115, 73H2200294, 73H2200433, 73H2200453, 73J2200747, 73J2200763, 73K2200014, 73K2200095, 73K2200101, 73K2200299, 73K2200566, 73K2200713, 73L2200064, 73L2200066 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

worldwide

No US distribution state could be established from this record. Check the official notice.

## Dates in the source

- **Report date**: Jun 7, 2023
- **Recall initiation date**: Apr 28, 2023
- **Center classification date**: May 30, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.