---
title: "NeuRx Diaphragm Pacing System (NeuRx DPS) External Pulse Generator (EPG), Par... — FDA record Z-1561-2022 | Is It on Recall"
description: "Due to a defect identified on the EPG controller printed circuit board assembly, device performance may degrade over time. Depending on the failure mode, potent"
url: https://isitonrecall.com/recall/fda-device-z-1561-2022-neurx-diaphragm-pacing-system-neurx-dps-external-pulse-generator-epg-part-no-23-0021/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# NeuRx Diaphragm Pacing System (NeuRx DPS) External Pulse Generator (EPG), Part No. 23-0021. — FDA record Z-1561-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 11, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: NeuRx Diaphragm Pacing System (NeuRx DPS) External Pulse Generator (EPG), Part No. 23-0021.
- **Reason**: Due to a defect identified on the EPG controller printed circuit board assembly, device performance may degrade over time. Depending on the failure mode, potential adverse events include pain, muscle fatigue and decreased breathing effectiveness, arrhythmia or cardiac arrest, or the device may stop unexpectedly.
- **Classification**: Class I
- **Recalling firm**: Synapse Biomedical Inc
- **Quantity**: 118 (112 US, 6 OUS)
- **Agency-reported status**: Terminated
- **Source record identifier**: Z-1561-2022

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1561-2022%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | UDI-DI: 00852184003212; Serial Nos. 1900 thru 2017 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

distribution to US states of AL, CO, IL, FL, GA, MD, MN, NC, NJ, OH, PA, and SC; Germany and Kuwait. \*\*\*Updated 3/28/23 - Product was not distributed to United Arab Emirates. Product was distributed to Kuwait.\*\*\* Updated 9.14.23 - 1 patient in Germany, 1 patient in the Czech Republic, 1 distributor in Kuwait

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Alabama](https://isitonrecall.com/state/alabama/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Georgia](https://isitonrecall.com/state/georgia/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Maryland](https://isitonrecall.com/state/maryland/)
- [Minnesota](https://isitonrecall.com/state/minnesota/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [Ohio](https://isitonrecall.com/state/ohio/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [South Carolina](https://isitonrecall.com/state/south-carolina/)

## Dates in the source

- **Report date**: Aug 31, 2022
- **Recall initiation date**: Jul 11, 2022
- **Center classification date**: Aug 22, 2022
- **Termination date**: Oct 24, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1561-2022%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Synapse Biomedical Inc records](https://isitonrecall.com/brand/synapse-biomedical-inc/)
