---
title: "Hillrom Welch Allyn Spot Vision Screener VS100, marketed under the following ... — FDA record Z-1510-2024 | Is It on Recall"
description: "Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and intern"
url: https://isitonrecall.com/recall/fda-device-z-1510-2024-hillrom-welch-allyn-spot-vision-screener-vs100-marketed-under-the-following-models-a-spot-vision-scr/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Hillrom Welch Allyn Spot Vision Screener VS100, marketed under the following models: a) SPOT VISION SCREENER,W/O CASE... — FDA record Z-1510-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 26, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Hillrom Welch Allyn Spot Vision Screener VS100, marketed under the following models: a) SPOT VISION SCREENER,W/O CASE,PLUG2/EUR, VS100-2; b) SPOT VISION SCREENER,W/O CASE,PLUG4/UK, VS100-4; c) SPOT VISION SCREENER,W/O CASE,PLUG7/SA, VS100-7; d) SPOT VISION SCREENER,W/CASE,PLUG2/EUR, VS100S-2; e) SPOT VISION SCREENER,W/CASE,PLUG4/UK, VS100S-4; f) SPOT VISION SCREENER,W/CASE,PLUG7/SA, VS100S-7; g) SPOT VISION SCREENER W/CASE, BRAZIL, VS100S-Z; h) SPOT VS POWER SUPPLY SET/PLUG2/EUR, 106364; i) SPOT VS POWER SUPPLY SET/PLUG4/UK, 106366; j) SPOT VS POWER SUPPLY SET/PLUG7/SA, 106368; for use during vision screening
- **Reason**: Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and international electrical standards.
- **Classification**: Class II
- **Recalling firm**: Baxter Healthcare Corporation
- **Quantity**: 3547 units
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-1510-2024

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1510-2024%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | a) VS100-2, UDI/DI 732094224214; b) VS100-4, UDI/DI 732094224191; c) VS100-7, UDI/DI 732094224092; d) VS100S-2, UDI/DI 732094224177; e) VS100S-4, UDI/DI 732094224153; f) VS100S-7, UDI/DI 732094224085; g) VS100S-Z, UDI/DI 732094294217; h) 106364, UDI/DI 732094246964; i) 106366, UDI/DI 732094319927; j) 106368, UDI/DI 732094248692 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Global, US: NY, IL

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Illinois](https://isitonrecall.com/state/illinois/)
- [New York](https://isitonrecall.com/state/new-york/)

## Dates in the source

- **Report date**: Apr 17, 2024
- **Recall initiation date**: Mar 26, 2024
- **Center classification date**: Apr 10, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1510-2024%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Baxter Healthcare Corporation records](https://isitonrecall.com/brand/baxter-healthcare-corporation/)
