---
title: "Hillrom Welch Allyn Connex ProBP 3400 Digital Blood Pressure Device, marketed... — FDA record Z-1509-2024 | Is It on Recall"
description: "Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and intern"
url: https://isitonrecall.com/recall/fda-device-z-1509-2024-hillrom-welch-allyn-connex-probp-3400-digital-blood-pressure-device-marketed-under-the-following-mod/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Hillrom Welch Allyn Connex ProBP 3400 Digital Blood Pressure Device, marketed under the following models: a) PROBP 34... — FDA record Z-1509-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 26, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Hillrom Welch Allyn Connex ProBP 3400 Digital Blood Pressure Device, marketed under the following models: a) PROBP 3400, SUREBP, BLUETOOTH, MOBILE,EU, Product Code 34BFST-2; b) PROBP 3400, SUREBP, BLUETOOTH, MOBILE,UK, Product Code 34BFST-4; c) PROBP 3400, SUREBP, BLUETOOTH, MOBILE,AU, Product Code 34BFST-6; d) PROBP 3400, BLUETOOTH, MOBILE, EU, Product Code 34BXST-2; e) PROBP 3400, BLUETOOTH, MOBILE, UK, Product Code 34BXST-4; f) PROBP 3400, BLUETOOTH, MOBILE, ZA, Product Code 34BXST-7; g) PROBP 3400, SUREBP, MOBILE, EU, Product Code 34XFST-2; h) PROBP 3400, SUREBP, MOBILE, UK, Product Code 34XFST-4; i) PROBP 3400, SUREBP, MOBILE, AU, Product Code 34XFST-6; j) PROBP 3400, SUREBP, MOBILE, ZA, Product Code 34XFST-7; k) PROBP 3400, MOBILE, EU, Product Code 34XXST-2; l) PROBP 3400, MOBILE, UK, Product Code 34XXST-4; m) PROBP 3400, MOBILE, AU, Product Code 34XXST-6; n) PROBP 3400, MOBILE, ZA, Product Code 34XXST-7; o) PROBP MOBILE STAND CORD ASSY EUROPE, Product Code PWCD-5WT-2; p) PROBP MOBILE STAND CORD ASSY UK, Product Code PWCD-5WT-4; q) PROBP MOBILE STAND CORD ASSY AUST, Product Code PWCD-5WT-6; r) PROBP MOBILE STAND CORD ASSY S. AFRICA, Product Code PWCD-5WT-7; s) PROBP MOBILE STAND CORD ASSY EUROPE, Product Code PWCD-6WT-2; t) PROBP MOBILE STAND CORD ASSY AUST, Product Code PWCD-6WT-6; u) PROBP MOBILE STAND CORD ASSY S. AFRICA, Product Code PWCD-6WT-7
- **Reason**: Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and international electrical standards.
- **Classification**: Class II
- **Recalling firm**: Baxter Healthcare Corporation
- **Quantity**: 11154 units
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-1509-2024

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1509-2024%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | a) 34BFST-2, UDI/DI 732094009385; b) 34BFST-4, UDI/DI 732094009378; c) 34BFST-6, UDI/DI 732094009361; d) 34BXST-2, UDI/DI 732094009248; e) 34BXST-4, UDI/DI 732094167733; f) 34BXST-7, UDI/DI 732094009231; g) 34XFST-2, UDI/DI 732094009057; h) 34XFST-4, UDI/DI 732094009040; i) 34XFST-6, UDI/DI 732094009033; j) 34XFST-7, UDI/DI 732094009026; k) 34XXST-2, UDI/DI 732094008852; l) 34XXST-4, UDI/DI 732094008845; m) 34XXST-6, UDI/DI 732094008838; n) 34XXST-7, UDI/DI 732094008821; o) PWCD-5WT-2, UDI/DI 732094072402; p) PWCD-5WT-4, UDI/DI 732094155808; q) PWCD-5WT-6, UDI/DI 732094072389; r) PWCD-5WT-7, UDI/DI 732094072365; s) PWCD-6WT-2, UDI/DI 732094347494; t) PWCD-6WT-6, UDI/DI 732094347470; u) PWCD-6WT-7, UDI/DI 732094347463 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Global, US: NY, IL

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Illinois](https://isitonrecall.com/state/illinois/)
- [New York](https://isitonrecall.com/state/new-york/)

## Dates in the source

- **Report date**: Apr 17, 2024
- **Recall initiation date**: Mar 26, 2024
- **Center classification date**: Apr 10, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1509-2024%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Baxter Healthcare Corporation records](https://isitonrecall.com/brand/baxter-healthcare-corporation/)
