title: "Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 ... — FDA record Z-1249-2025 | Is It on Recall"
description: "Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the aff"
url: https://isitonrecall.com/recall/fda-device-z-1249-2025-access-erythropoietin-epo-catalog-number-a16364-a-glycoprotein-30-400-daltons-produced-primarily-by/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidne... — FDA record Z-1249-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 3, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal factor regulating red blood cell production (erythropoiesis) in mammals. Renal production of EPO is regulated by changes in oxygen availability. Under conditions of hypoxia, the level of EPO in the circulation increases and this leads to increased production of red blood cells.
- **Reason**: Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased results, which may lead to improper diagnosis or repeat testing.
- **Classification**: Class II
- **Recalling firm**: Beckman Coulter, Inc.
- **Quantity**: 498 US; 1576 OUS
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-1249-2025

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1249-2025%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Catalog A16364 UDI-DI 15099590201838 Lot 439363 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Domestic distribution to AL AR AZ CA CT FL ID IL IN KS MA MI NC NE NJ NY OH OR PA TN TX UT WA WV. International distribution to Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Korea, Republic of, Netherlands, Philippines, Russian Federation, Saudi Arabia, Slovakia, Spain, United Kingdom of Great Britain and Northern Ireland

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Alabama](https://isitonrecall.com/state/alabama/)
- [Arkansas](https://isitonrecall.com/state/arkansas/)
- [Arizona](https://isitonrecall.com/state/arizona/)
- [California](https://isitonrecall.com/state/california/)
- [Connecticut](https://isitonrecall.com/state/connecticut/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Idaho](https://isitonrecall.com/state/idaho/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Kansas](https://isitonrecall.com/state/kansas/)
- [Massachusetts](https://isitonrecall.com/state/massachusetts/)
- [Michigan](https://isitonrecall.com/state/michigan/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [Nebraska](https://isitonrecall.com/state/nebraska/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Ohio](https://isitonrecall.com/state/ohio/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [Tennessee](https://isitonrecall.com/state/tennessee/)
- [Texas](https://isitonrecall.com/state/texas/)
- [Utah](https://isitonrecall.com/state/utah/)
- [Washington](https://isitonrecall.com/state/washington/)
- [West Virginia](https://isitonrecall.com/state/west-virginia/)

## Dates in the source

- **Report date**: Mar 12, 2025
- **Recall initiation date**: Feb 3, 2025
- **Center classification date**: Feb 28, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1249-2025%22)

## Related records